Abstract
Objective: To ascertain if wideband acoustic immitance (WAI) measurements are useful in assessing crimping status following stapedotomy. Design: WAI measurements were obtained using the Mimosa Acoustics HearID system. Wideband chirp sound stimuli and a set of tone stimuli for nine frequencies between 0.2 and 6 kHz were used at 60 dB SPL. Five sets of measurements were performed on each temporal bone: mobile stapes, stapes fixation and stapedotomy followed by insertion of a tightly crimped, a loosely crimped and an uncrimped prosthesis. Study sample: Eight fresh-frozen temporal bones were harvested from human cadaveric donors. Results: At lower frequencies, up to 1 kHz, stapes fixation decreased absorbance. Compared to the baseline absorbance, absorbance with stapes fixation dropped by 6 to 17% in absolute terms from the baseline value (p = 0.027). Absorbance was not affected in higher frequencies (p = 0.725). Stapedotomy changed the absorbance curve significantly compared to the normal condition with an increase of absolute absorbance values by 6 to 36% around 0.25–1 kHz (p-value
| Original language | English |
|---|---|
| Pages (from-to) | 1-7 |
| Number of pages | 7 |
| Journal | International Journal of Audiology |
| Volume | 56 |
| Issue number | 1 |
| DOIs | |
| Publication status | Published - 2017 |
Keywords
- Otosclerosis
- stapes
- surgery
- stapedotomy
- middle ear
- temporal bone
- wideband
- absorbance
- power absorption
- immitance
- impedance
- acoustic resistance
- transmittance
- reflectance
- hearing loss
Fingerprint
Dive into the research topics of 'Wideband acoustic immittance measurements in assessing crimping status following stapedotomy: A temporal bone study'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver