BACKGROUND: Patients undergoing carotid endarterectomy require strict arterial blood pressure (BP) control to maintain adequate cerebral perfusion. In this study we tested whether non-invasive beat-to-beat Nexfin finger BP (BPfin) can replace invasive beat-to-beat radial artery BP (BPrad) in this setting.

METHODS: In 25 consecutive patients (median age 71 yr) scheduled for carotid endarterectomy and receiving general anaesthesia, BPfin and BPrad were monitored simultaneously and ipsilaterally during the 30-min period surrounding carotid artery cross-clamping. Validation was guided by the standard set by the Association for the Advancement of Medical Instrumentation (AAMI), which considers a BP monitor adequate when bias (precision) is <5 (8) mm Hg, respectively.

RESULTS: BPfin vs BPrad bias (precision) was -3.3 (10.8), 6.1 (5.7) and 3.5 (5.2) mm Hg for systolic, diastolic, and mean BP, respectively. One subject was excluded due to a poor quality BP curve. In another subject, mean BPfin overestimated mean BPrad by 13.5 mm Hg.

CONCLUSION: Mean BPfin could be considered as an alternative for mean BPrad during a carotid endarterectomy, based on the AAMI criteria. In 23 of 24 patients, the use of mean BPfin would not lead to decisions to adjust mean BPrad values outside the predefined BP threshold.

CLINICALTRIALSGOV: NCT01451294.

Original languageEnglish
Pages (from-to)316-323
Number of pages8
JournalBritish Journal of Anaesthesia
Volume117
Issue number3
DOIs
Publication statusPublished - 1 Sept 2016

Keywords

  • carotid endarterectomy
  • measurement techniques, arterial pressure
  • Nexfin
  • non-invasive continuous arterial pressure measurement

Fingerprint

Dive into the research topics of 'Validation of non-invasive arterial pressure monitoring during carotid endarterectomy'. Together they form a unique fingerprint.

Cite this