TY - JOUR
T1 - Validation of digital microscopy: Review of validation methods and sources of bias
AU - Bertram, Christof A.
AU - Stathonikos, Nikolas
AU - Donovan, Taryn A.
AU - Bartel, Alexander
AU - Fuchs-Baumgartinger, Andrea
AU - Lipnik, Karoline
AU - van Diest, Paul J.
AU - Bonsembiante, Federico
AU - Klopfleisch, Robert
N1 - Funding Information:
The author(s) disclosed receipt of the following financial support for the research, authorship, and/or publication of this article: C.A. Bertram gratefully acknowledges the scholarship from the Dres. Jutta & Georg Bruns Stiftung für innovative Veterinärmedizin.
Publisher Copyright:
© The Author(s) 2021.
PY - 2022/1/1
Y1 - 2022/1/1
N2 - Digital microscopy (DM) is increasingly replacing traditional light microscopy (LM) for performing routine diagnostic and research work in human and veterinary pathology. The DM workflow encompasses specimen preparation, whole-slide image acquisition, slide retrieval, and the workstation, each of which has the potential (depending on the technical parameters) to introduce limitations and artifacts into microscopic examination by pathologists. Performing validation studies according to guidelines established in human pathology ensures that the best-practice approaches for patient care are not deteriorated by implementing DM. Whereas current publications on validation studies suggest an overall high reliability of DM, each laboratory is encouraged to perform an individual validation study to ensure that the DM workflow performs as expected in the respective clinical or research environment. With the exception of validation guidelines developed by the College of American Pathologists in 2013 and its update in 2021, there is no current review of the application of methods fundamental to validation. We highlight that there is high methodological variation between published validation studies, each having advantages and limitations. The diagnostic concordance rate between DM and LM is the most relevant outcome measure, which is influenced (regardless of the viewing modality used) by different sources of bias including complexity of the cases examined, diagnostic experience of the study pathologists, and case recall. Here, we review 3 general study designs used for previous publications on DM validation as well as different approaches for avoiding bias.
AB - Digital microscopy (DM) is increasingly replacing traditional light microscopy (LM) for performing routine diagnostic and research work in human and veterinary pathology. The DM workflow encompasses specimen preparation, whole-slide image acquisition, slide retrieval, and the workstation, each of which has the potential (depending on the technical parameters) to introduce limitations and artifacts into microscopic examination by pathologists. Performing validation studies according to guidelines established in human pathology ensures that the best-practice approaches for patient care are not deteriorated by implementing DM. Whereas current publications on validation studies suggest an overall high reliability of DM, each laboratory is encouraged to perform an individual validation study to ensure that the DM workflow performs as expected in the respective clinical or research environment. With the exception of validation guidelines developed by the College of American Pathologists in 2013 and its update in 2021, there is no current review of the application of methods fundamental to validation. We highlight that there is high methodological variation between published validation studies, each having advantages and limitations. The diagnostic concordance rate between DM and LM is the most relevant outcome measure, which is influenced (regardless of the viewing modality used) by different sources of bias including complexity of the cases examined, diagnostic experience of the study pathologists, and case recall. Here, we review 3 general study designs used for previous publications on DM validation as well as different approaches for avoiding bias.
KW - accuracy
KW - concordance rate
KW - digital microscopy
KW - digital pathology
KW - noninferiority
KW - review
KW - study design
KW - validation
KW - virtual microscopy
KW - whole-slide images
UR - http://www.scopus.com/inward/record.url?scp=85113616066&partnerID=8YFLogxK
U2 - 10.1177/03009858211040476
DO - 10.1177/03009858211040476
M3 - Review article
C2 - 34433345
SN - 0300-9858
VL - 59
SP - 26
EP - 38
JO - Veterinary Pathology
JF - Veterinary Pathology