T1 - Validation of automated positive cell and region detection of immunohistochemically stained laryngeal tumor tissue using digital image analysis
AU - Smits, Hilde J.G.
AU - Swartz, Justin E.
AU - Philippens, Marielle E.P.
AU - de Bree, Remco
AU - Kaanders, Johannes H.A.M.
AU - Koppes, Sjors A.
AU - Breimer, Gerben E.
AU - Willems, Stefan M.
N1 - Funding Information:
The original Phase III trial was supported by the Dutch Cancer Society (KWF) Research Fund No. CKTO-2000-09 and KUN-2008-4088 and a European Community Grant No. FP7-HEALTH-2007-B222741 , project METOXIA (Metastatic Tumors Facilitated by Hypoxic Tumor Micro-Environments) . No additional funding was received for the present study.
Publisher Copyright:
© 2023 The Authors
PY - 2023/1
Y1 - 2023/1
N2 - Objectives: This study aimed to validate a digital image analysis (DIA) workflow for automatic positive cell detection and positive region delineation for immunohistochemical hypoxia markers with a nuclear (hypoxia-inducible factor 1α [HIF-1α]) and a cytoplasmic (pimonidazole [PIMO]) staining pattern. Materials and methods: 101 tissue fragments from 44 laryngeal tumor biopsies were immunohistochemically stained for HIF-1α and PIMO. QuPath was used to determine the percentage of positive cells and to delineate positive regions automatically. For HIF-1α, only cells with strong staining were considered positive. Three dedicated head and neck pathologists scored the percentage of positive cells using three categories (0: <1%; 1: 1%–33%; 2: >33%;). The pathologists also delineated the positive regions on 14 corresponding PIMO and HIF-1α-stained fragments. The consensus between observers was used as the reference standard and was compared to the automatic delineation. Results: Agreement between categorical positivity scores was 76.2% and 65.4% for PIMO and HIF-1α, respectively. In all cases of disagreement in HIF-1α fragments, the DIA underestimated the percentage of positive cells. As for the region detection, the DIA correctly detected most positive regions on PIMO fragments (false positive area=3.1%, false negative area=0.7%). In HIF-1α, the DIA missed some positive regions (false positive area=1.3%, false negative area=9.7%). Conclusions: Positive cell and region detection on biopsy material is feasible, but further optimization is needed before unsupervised use. Validation at varying DAB staining intensities is hampered by lack of reliability of the gold standard (i.e., visual human interpretation). Nevertheless, the DIA method has the potential to be used as a tool to assist pathologists in the analysis of IHC staining.
AB - Objectives: This study aimed to validate a digital image analysis (DIA) workflow for automatic positive cell detection and positive region delineation for immunohistochemical hypoxia markers with a nuclear (hypoxia-inducible factor 1α [HIF-1α]) and a cytoplasmic (pimonidazole [PIMO]) staining pattern. Materials and methods: 101 tissue fragments from 44 laryngeal tumor biopsies were immunohistochemically stained for HIF-1α and PIMO. QuPath was used to determine the percentage of positive cells and to delineate positive regions automatically. For HIF-1α, only cells with strong staining were considered positive. Three dedicated head and neck pathologists scored the percentage of positive cells using three categories (0: <1%; 1: 1%–33%; 2: >33%;). The pathologists also delineated the positive regions on 14 corresponding PIMO and HIF-1α-stained fragments. The consensus between observers was used as the reference standard and was compared to the automatic delineation. Results: Agreement between categorical positivity scores was 76.2% and 65.4% for PIMO and HIF-1α, respectively. In all cases of disagreement in HIF-1α fragments, the DIA underestimated the percentage of positive cells. As for the region detection, the DIA correctly detected most positive regions on PIMO fragments (false positive area=3.1%, false negative area=0.7%). In HIF-1α, the DIA missed some positive regions (false positive area=1.3%, false negative area=9.7%). Conclusions: Positive cell and region detection on biopsy material is feasible, but further optimization is needed before unsupervised use. Validation at varying DAB staining intensities is hampered by lack of reliability of the gold standard (i.e., visual human interpretation). Nevertheless, the DIA method has the potential to be used as a tool to assist pathologists in the analysis of IHC staining.
KW - Biomarker analysis
KW - Computational pathology
KW - Positive cell detection
KW - Positive region detection
KW - Validation
UR - http://www.scopus.com/inward/record.url?scp=85148712558&partnerID=8YFLogxK
U2 - 10.1016/j.jpi.2023.100198
DO - 10.1016/j.jpi.2023.100198
M3 - Article
C2 - 36818021
SN - 2229-5089
VL - 14
JO - Journal of Pathology Informatics
JF - Journal of Pathology Informatics
M1 - 100198
ER -