Abstract

BACKGROUND: The new balloon-expandable Sapien 3 transcatheter heart valve (S3-THV) incorporates new features to reduce aortic regurgitation (AR) and vascular complications in transcatheter aortic valve implantation. Aim of this study is to compare the outcomes of the S3-THV with the preceding Sapien XT THV (SXT-THV) in patients who underwent transcatheter aortic valve implantation for symptomatic severe native aortic stenosis.

METHODS AND RESULTS: Eligible patients were retrospectively identified in our institutional database and periprocedural clinical and imaging data were collected. Non-parsimonious one-to-many propensity score matching was performed to account for differences in baseline characteristics. Between November 2011 and December 2014, 167 patients underwent balloon-expandable transcatheter aortic valve implantation with either the S3-THV (n=49) or SXT-THV (n=118). Forty-four (89.8%) S3-THV patients were matched to 66 (55.9%) SXT-THV patients (mean age 80.3±8.4 and 80.5±7.8 years, median EuroSCORE 15.8 and 16.5%, respectively). In the S3-THV and SXT-THV groups, transfemoral approach (77.3% versus 78.8%) and postdilatation rates (15.9% versus 12.1%) were similar. Predischarge echocardiography demonstrated a lower incidence of ≥mild AR (15.9% versus 46.2%, P=0.003) for the S3-THV, despite reduced annulus area to prosthesis oversizing (8.2±5.1 versus 18.2±10.7%, P=0.001). Transfemoral access site-related life-threatening or major bleedings and vascular complications were absent in the S3-THV group (0% versus 7.7%, P=0.15). No differences were observed in pacemaker implantation rate (9.8% versus 8.8%, P=0.94) and 30-day mortality (both 5%).

CONCLUSIONS: In this retrospective, propensity score-matched analysis, the S3-THV performed superiorly to the SXT-THV, as demonstrated by improved valve patency and increased transfemoral access safety.

Original languageEnglish
Article numbere002408
Number of pages20
JournalCirculation. Cardiovascular Interventions
Volume8
Issue number6
DOIs
Publication statusPublished - 1 Jan 2015

Keywords

  • Aortic regurgitation
  • aortic valve stenosis
  • transcatheter aortic valve implantation
  • vascular complications

Fingerprint

Dive into the research topics of 'Transcatheter aortic valve implantation with the new balloon-expandable sapien 3 versus sapien XT valve system'. Together they form a unique fingerprint.

Cite this