Towards harmonized holmium-166 SPECT image quality for dosimetry: a multi-center, multi-vendor study

Lovisa E.L. Westlund Gotby*, Martina Stella, Camille D.E. Van Speybroeck, Daphne Lobeek, Floris H.P. van Velden, Mette K. Stam, Petra Dibbets-Schneider, Daphne M.V. de Vries-Huizing, Erik Jan Rijkhorst, Berlinda J. de Wit-van de Veen, Roel Wierts, Rob van Rooij

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

1 Downloads (Pure)

Abstract

Background: Reliable dosimetry based on SPECT/CT imaging is essential to achieve personalized 166Ho-radioembolization treatment planning and evaluation. This study quantitatively evaluates multiple acquisition and reconstruction protocols for 166Ho-SPECT imaging based on data from five Dutch hospitals. We aim to recommend an imaging protocol which harmonizes 166Ho-SPECT images for reproducible and accurate dosimetry in a multi-scanner and multi-center setting. Methods: Cylindrical and NEMA IEC phantoms, filled with 166Ho-chloride, were imaged using seven SPECT/CT scanners from two vendors (GE HealthCare and Siemens Healthineers). Data were acquired with a photopeak window centered at 81 keV. Two adjacent scatter windows, and one upper scatter window at 118 keV were used for triple-energy window (TEW) and dual-energy window (DEW) scatter correction, respectively. The TEW and DEW reconstructions used vendor-specific software. Additionally, a vendor-neutral software package with Monte Carlo (MC) scatter correction (Hermes Medical Solutions) was used to study the influence of scanner hardware on the image quality. System sensitivity was measured in projection data of the cylindrical phantom. The axial uniformity in the cylindrical phantom was used to characterize the impact of the scatter correction method. The image quality was evaluated by the coefficient of variation (COV; noise), the contrast recovery coefficients (CRCs) and contrast-to-noise ratios (CNRs). Results: TEW scatter correction resulted in superior uniformity and higher CRCs compared to the DEW (CRC for the largest sphere over all scanners, mean ± SD (range): TEW 0.54 ± 0.07 (0.36–0.65), DEW 0.44 ± 0.04 (0.34–0.51)). DEW resulted in lower noise levels compared to TEW (16% lower on average). The DEW and TEW images resulted in comparable CNRs. The system sensitivities and the vendor-neutral image reconstructions demonstrated differences in hardware between the two vendors, most likely due to the characteristics of the vendor-specific medium energy collimator. Conclusion: This study demonstrates that TEW scatter correction increases the accuracy of 166Ho-SPECT images compared to DEW, and we henceforth recommend adopting this method in the clinical 166Ho-dosimetry workflow. Scanner hardware has a substantial impact on the characteristics of the acquired data, and identical reconstruction settings will therefore not automatically lead to harmonized image quality.

Original languageEnglish
Article number24
Number of pages19
JournalEJNMMI physics
Volume12
Issue number1
DOIs
Publication statusPublished - 19 Mar 2025

Keywords

  • Dosimetry
  • Dual-energy window scatter correction
  • Holmium-166
  • Image quality
  • Monte Carlo-based scatter correction
  • Phantom study
  • Radioembolization
  • SPECT/CT
  • Triple-energy window scatter correction

Fingerprint

Dive into the research topics of 'Towards harmonized holmium-166 SPECT image quality for dosimetry: a multi-center, multi-vendor study'. Together they form a unique fingerprint.

Cite this