Tier-based formalism for safety assessment of custom-built radio-frequency transmit coils

B R Steensma*, A Sadeghi-Tarakameh, E F Meliado, C A T van den Berg, D W J Klomp, P R Luijten, G J Metzger, Y Eryaman, A J E Raaijmakers

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

24 Downloads (Pure)

Abstract

The purpose of this work is to propose a tier-based formalism for safety assessment of custom-built radio-frequency (RF) coils that balances validation effort with the effort put in determinating the safety factor. The formalism has three tier levels. Higher tiers require increased effort when validating electromagnetic simulation results but allow for less conservative safety factors. In addition, we propose a new method to calculate modeling uncertainty between simulations and measurements and a new method to propagate uncertainties in the simulation into a safety factor that minimizes the risk of underestimating the peak specific absorption rate (SAR). The new safety assessment procedure was completed for all tier levels for an eight-channel dipole array for prostate imaging at 7 T and an eight-channel dipole array for head imaging at 10.5 T, using data from two different research sites. For the 7 T body array, the validation procedure resulted in a modeling uncertainty of 77% between measured and simulated local SAR distributions. For a situation where RF shimming is performed on the prostate, average power limits of 2.4 and 4.5 W/channel were found for tiers 2 and 3, respectively. When the worst-case peak SAR among all phase settings was calculated, power limits of 1.4 and 2.7 W/channel were found for tiers 2 and 3, respectively. For the 10.5 T head array, a modeling uncertainty of 21% was found based on B 1 + mapping. For the tier 2 validation, a power limit of 2.6 W/channel was calculated. The demonstrated tier system provides a strategy for evaluating modeling inaccuracy, allowing for the rapid translation of novel coil designs with conservative safety factors and the implementation of less conservative safety factors for frequently used coil arrays at the expense of increased validation effort.

Original languageEnglish
Article numbere4874
Number of pages16
JournalNMR in Biomedicine
Volume36
Issue number5
Early online date11 Nov 2022
DOIs
Publication statusPublished - May 2023

Keywords

  • EM simulations
  • RF coils
  • safety factor
  • SAR
  • validation

Fingerprint

Dive into the research topics of 'Tier-based formalism for safety assessment of custom-built radio-frequency transmit coils'. Together they form a unique fingerprint.