TY - JOUR
T1 - Thermal ablation of a confluent lesion in the porcine kidney with a clinically available MR-HIFU system
AU - van Breugel, Johanna Maria Mijntje
AU - de Greef, Martijn
AU - Wijlemans, JW
AU - Schubert, Gerald
AU - van den Bosch, Maurice
AU - Moonen, Chrit
AU - Ries, Mario
N1 - Publisher Copyright:
© 2017 Institute of Physics and Engineering in Medicine.
PY - 2017
Y1 - 2017
N2 - The incidence of small renal masses (SRMs) sized <4 cm has increased over the decades (as co-findings/or due to introduction of cross sectional imaging). Currently, partial nephrectomy (PN) or watchful waiting is advised in these patients. Ultimately, 80-90% of these SRMs require surgical treatment and PN is associated with a 15% complication rate. In this aging population, with possible comorbidities and poor health condition, both PN and watchful waiting are non-ideal treatment options. This resulted in an increased need for early, non-invasive treatment strategies such as MR-guided high intensity focused ultrasound (MR-HIFU). (i) To investigate the feasibility of creating a confluent lesion in the kidney using respiratory-gated MR-HIFU under clinical conditions in a pre-clinical study and (ii) to evaluate the reproducibility of the MR-HIFU ablation strategy. Healthy pigs (n = 10) under general anesthesia were positioned on a clinical MR-HIFU system with integrated cooling. A honeycomb pattern of seven overlapping ablation cells (4 × 4 × 10 mm
3, 450 W, <30 s) was ablated successively in the cortex of the porcine kidney. Both MR thermometry and acoustic energy delivery were respiratory gated using a pencil beam navigator on the contralateral kidney. The non-perfused volume (NPV) was visualized after the last sonication by contrast-enhanced (CE) T
1-weighted MR (T
1 w) imaging. Cell viability staining was performed to visualize the extent of necrosis. Results: a median NPV of 0.62 ml was observed on CE-T
1 w images (IQR 0.58-1.57 ml, range 0.33-2.75 ml). Cell viability staining showed a median damaged volume of 0.59 ml (IQR 0.24-1.35 ml, range 0-4.1 ml). Overlooking of the false rib, shivering of the pig, and too large depth combined with a large heat-sink effect resulted in insufficient heating in 4 cases. The NPV and necrosed volume were confluent in all cases in which an ablated volume could be observed. Our results demonstrated the feasibility of creating a confluent volume of ablated kidney cortical tissue in vivo with MR-HIFU on a clinically available system using respiratory gating and near-field cooling and showed its reproducibility.
AB - The incidence of small renal masses (SRMs) sized <4 cm has increased over the decades (as co-findings/or due to introduction of cross sectional imaging). Currently, partial nephrectomy (PN) or watchful waiting is advised in these patients. Ultimately, 80-90% of these SRMs require surgical treatment and PN is associated with a 15% complication rate. In this aging population, with possible comorbidities and poor health condition, both PN and watchful waiting are non-ideal treatment options. This resulted in an increased need for early, non-invasive treatment strategies such as MR-guided high intensity focused ultrasound (MR-HIFU). (i) To investigate the feasibility of creating a confluent lesion in the kidney using respiratory-gated MR-HIFU under clinical conditions in a pre-clinical study and (ii) to evaluate the reproducibility of the MR-HIFU ablation strategy. Healthy pigs (n = 10) under general anesthesia were positioned on a clinical MR-HIFU system with integrated cooling. A honeycomb pattern of seven overlapping ablation cells (4 × 4 × 10 mm
3, 450 W, <30 s) was ablated successively in the cortex of the porcine kidney. Both MR thermometry and acoustic energy delivery were respiratory gated using a pencil beam navigator on the contralateral kidney. The non-perfused volume (NPV) was visualized after the last sonication by contrast-enhanced (CE) T
1-weighted MR (T
1 w) imaging. Cell viability staining was performed to visualize the extent of necrosis. Results: a median NPV of 0.62 ml was observed on CE-T
1 w images (IQR 0.58-1.57 ml, range 0.33-2.75 ml). Cell viability staining showed a median damaged volume of 0.59 ml (IQR 0.24-1.35 ml, range 0-4.1 ml). Overlooking of the false rib, shivering of the pig, and too large depth combined with a large heat-sink effect resulted in insufficient heating in 4 cases. The NPV and necrosed volume were confluent in all cases in which an ablated volume could be observed. Our results demonstrated the feasibility of creating a confluent volume of ablated kidney cortical tissue in vivo with MR-HIFU on a clinically available system using respiratory gating and near-field cooling and showed its reproducibility.
KW - clinical MR-HIFU
KW - confluent lesion
KW - kidney
KW - pencil beam navigator
KW - renal lesion
KW - respiratory gating
KW - thermal ablation
UR - http://www.scopus.com/inward/record.url?scp=85020494067&partnerID=8YFLogxK
U2 - 10.1088/1361-6560/aa75b3
DO - 10.1088/1361-6560/aa75b3
M3 - Article
C2 - 28557798
SN - 0031-9155
VL - 62
SP - 5312
EP - 5326
JO - Physics in Medicine and Biology
JF - Physics in Medicine and Biology
IS - 13
ER -