T1 - The effect of catheter displacement and anatomical variations on the dose distribution in MRI-guided focal HDR brachytherapy for prostate cancer
AU - Maenhout, Metha
AU - van der Voort van Zyp, Jochem R N
AU - Borot, Maxence
AU - Peters, Max
AU - van Vulpen, Marco
AU - van den Bosch, Maurice
AU - Moerland, Rien
N1 - Copyright © 2017 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.
PY - 2018/1
Y1 - 2018/1
N2 - PURPOSE: The aim of this study was to analyze the effect of catheter displacement and anatomical variations of prostate and organs at risk on dose distribution in MRI-guided 19 Gy single fraction focal high-dose-rate brachytherapy (HDR-BT) of the prostate.METHODS AND MATERIALS: Seventeen patients with localized prostate cancer were enrolled in a prospective trial investigating focal HDR-BT in a 1.5 T MRI-HDR-BT facility. The diagnostic MRI delineations were registered with intraoperative MR scan, and a single fraction of 19 Gy was applied to the visible tumor. Self-anchoring umbrella catheters were used for HDR-BT delivery. A 1.5 T MRI was performed directly after ultrasound (US)-guided catheter placement for treatment planning. After treatment and before removal of catheters, a posttreatment 1.5 T MRI was performed. Regions of interest were also delineated on the posttreatment MR images and the catheters of 17 patients were reconstructed. The dose plan was constructed for the posttreatment MRI scan to assess the influence of catheter migration and anatomical variation on the dose delivered to the target and the organs at risk. Also on the posttreatment MRI, the complete catheter reconstruction was reassessed, to correct for, for example, bending of the catheters. The displacement of catheters between the MRI scans was determined by comparing the catheter tip positions on the treatment planning and posttreatment 1.5 T MRI scans.RESULTS: The displacements of 241 catheters were investigated. Average (range) displacements of the umbrella catheters are 0.6 (0-2.9) mm in the x-direction, 0.5 (0-2.1) mm in the y-direction, and 0.9 (0-5.5) mm in the z-direction. In 3 patients, the displacement was >4 mm and up to 5.5 mm. This occurred in respectively 1/13, 1/16, and 1/18 catheters in these patients. The dosimetric differences between the intraoperative treatment and the posttreatment plans were in most patients less than 1.5 Gy. In 4 patients, a dose difference in clinical target volume D95 of >2 Gy up to 5.8 Gy was reported. No discrimination can be made between dose differences due to catheter displacement and/or organ movement/anatomy changes.CONCLUSIONS: In general, catheter displacements were in the order of a mm and differences in dose to the clinical target volume and the organs at risk between the treatment and posttreatment plans smaller than 1.5 Gy. In some patients, dose differences up to 5.8 Gy were determined, due to either individual larger catheter displacement and/or anatomy changes. A longer followup is necessary to assess the clinical implications of individual large dose differences.
AB - PURPOSE: The aim of this study was to analyze the effect of catheter displacement and anatomical variations of prostate and organs at risk on dose distribution in MRI-guided 19 Gy single fraction focal high-dose-rate brachytherapy (HDR-BT) of the prostate.METHODS AND MATERIALS: Seventeen patients with localized prostate cancer were enrolled in a prospective trial investigating focal HDR-BT in a 1.5 T MRI-HDR-BT facility. The diagnostic MRI delineations were registered with intraoperative MR scan, and a single fraction of 19 Gy was applied to the visible tumor. Self-anchoring umbrella catheters were used for HDR-BT delivery. A 1.5 T MRI was performed directly after ultrasound (US)-guided catheter placement for treatment planning. After treatment and before removal of catheters, a posttreatment 1.5 T MRI was performed. Regions of interest were also delineated on the posttreatment MR images and the catheters of 17 patients were reconstructed. The dose plan was constructed for the posttreatment MRI scan to assess the influence of catheter migration and anatomical variation on the dose delivered to the target and the organs at risk. Also on the posttreatment MRI, the complete catheter reconstruction was reassessed, to correct for, for example, bending of the catheters. The displacement of catheters between the MRI scans was determined by comparing the catheter tip positions on the treatment planning and posttreatment 1.5 T MRI scans.RESULTS: The displacements of 241 catheters were investigated. Average (range) displacements of the umbrella catheters are 0.6 (0-2.9) mm in the x-direction, 0.5 (0-2.1) mm in the y-direction, and 0.9 (0-5.5) mm in the z-direction. In 3 patients, the displacement was >4 mm and up to 5.5 mm. This occurred in respectively 1/13, 1/16, and 1/18 catheters in these patients. The dosimetric differences between the intraoperative treatment and the posttreatment plans were in most patients less than 1.5 Gy. In 4 patients, a dose difference in clinical target volume D95 of >2 Gy up to 5.8 Gy was reported. No discrimination can be made between dose differences due to catheter displacement and/or organ movement/anatomy changes.CONCLUSIONS: In general, catheter displacements were in the order of a mm and differences in dose to the clinical target volume and the organs at risk between the treatment and posttreatment plans smaller than 1.5 Gy. In some patients, dose differences up to 5.8 Gy were determined, due to either individual larger catheter displacement and/or anatomy changes. A longer followup is necessary to assess the clinical implications of individual large dose differences.
KW - High-dose-rate
KW - Brachytherapy
KW - Prostate cancer
KW - Catheter displacement
KW - MRI guided
KW - Radiotherapy Dosage
KW - Prostate/diagnostic imaging
KW - Prospective Studies
KW - Humans
KW - Brachytherapy/methods
KW - Male
KW - Prostatic Neoplasms/diagnostic imaging
KW - Foreign-Body Migration/diagnostic imaging
KW - Catheters
KW - Organs at Risk/anatomy & histology
KW - Magnetic Resonance Imaging
KW - Ultrasonography
KW - Aged
KW - Radiotherapy Planning, Computer-Assisted
UR - http://www.scopus.com/inward/record.url?scp=85020121514&partnerID=8YFLogxK
U2 - 10.1016/j.brachy.2017.04.239
DO - 10.1016/j.brachy.2017.04.239
M3 - Article
C2 - 28579419
SN - 1538-4721
VL - 17
SP - 68
EP - 77
JO - Brachytherapy
JF - Brachytherapy
IS - 1
ER -