Abstract
OBJECTIVE: Little is known about the prevalence and degree of deformation of surgically implanted aortic biological valve prostheses (bio-sAVRs). We assessed bio-sAVR deformation using multidetector-row computed tomography (MDCT).
METHODS: Three imaging databases were searched for patients with MDCT performed after bio-sAVR implantation. Minimal and maximal valve ring diameters were obtained in systole and/or diastole, depending on the acquired cardiac phase(s). The eccentricity index (EI) was calculated as a measure of deformation as (1 - (minimal diameter/maximal diameter)) × 100%. EI of < 5% was considered none or trivial deformation, 5-10% mild deformation, and > 10% non-circular. Indications for MDCT and implanted valve type were retrieved.
RESULTS: One hundred fifty-two scans of bio-sAVRs were included. One hundred seventeen measurements were performed in systole and 35 in diastole. None or trivial deformation (EI < 5%) was seen in 67/152 (44%) of patients. Mild deformation (EI 5-10%) was seen in 59/152 (39%) and non-circularity was found in 26/152 (17%) of cases. Overall, median EI was 5.5% (IQR 3.4-7.8). In 77 patients, both systolic and diastolic measurements were performed from the same scan. For these scans, the median EI was 6.5% (IQR 3.4-10.2) in systole and 5.1% (IQR3.1-7.6) in diastole, with a significant difference between both groups (p = 0.006).
CONCLUSIONS: Surgically implanted aortic biological valve prostheses show mild deformation in 39% of cases and were considered non-circular in 17% of studied valves.
KEY POINTS: • Deformation of surgically implanted aortic valve bioprostheses (bio-sAVRs) can be adequately assessed using MDCT. • Bio-sAVRs show at least mild deformation (eccentricity index > 5%) in 56% of studied cases and were considered non-circular (eccentricity index > 10%) in 17% of studied valves. • The higher deformity rate found in bio-sAVRs with (suspected) valve pathology could suggest that geometric deformity may play a role in leaflet malformation and thrombus formation similar to that of transcatheter heart valves.
Original language | English |
---|---|
Pages (from-to) | 2651-2657 |
Number of pages | 7 |
Journal | European Radiology |
Volume | 30 |
Issue number | 5 |
DOIs | |
Publication status | Published - 1 May 2020 |
Keywords
- Bioprosthesis
- Heart valve prosthesis
- Multidetector-row computed tomography