T1 - Stability and comparison of complete blood count parameters between capillary and venous blood samples
AU - Doeleman, Martijn J.H.
AU - Esseveld, Anouk
AU - Huisman, Albert
AU - de Roock, Sytze
AU - Tiel Groenestege, Wouter M.
N1 - Publisher Copyright:
© 2023 The Authors. International Journal of Laboratory Hematology published by John Wiley & Sons Ltd.
PY - 2023/10
Y1 - 2023/10
N2 - Introduction: This study assessed the comparability of complete blood count (CBC) parameters between capillary and venous samples, and extended previous research by examining the influence of different storage temperatures on CBC stability up to 7 days after sample collection. Methods: Venous and capillary blood samples were collected from 93 adult patients. Hemoglobin (Hb), hematocrit (Ht), mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), mean corpuscular hemoglobin concentration (MCHC), mean platelet volume (MPV), leukocytes, lymphocytes, basophils, eosinophils, erythrocytes, red cell distribution width (RDW), immature granulocytes (IG), immature reticulocyte fraction (IRF), monocytes, neutrophils, platelets, and reticulocytes were measured. Deming regression and mean relative differences between venous and capillary measurements were contrasted with desirable total allowable error (TEa). Stability was assessed in 20–27 venous blood samples stored at 4, 21–22, or 30°C, and analyzed at 0, 24, 48, 72, 96, 120, 144, and 168 h. Mean relative change with respect to baseline measurements was compared to the desirable TEa to determine acceptable stability. Results: Deming regression demonstrated strong linear correlations and acceptable variation between venous and capillary measurements. Erythrocytes, Hb, Ht, MCH, MCV, RDW, reticulocytes, and platelets showed acceptable stability for at least 96 h at 4°C. Mean relative change exceeded desirable TEa after 24 h at 30°C for all parameters, except erythrocytes, Hb, leukocytes, and MCH. Conclusion: Clinical laboratory specialists and clinicians should be aware of potential differences between venous and capillary measurements, and the influence of storage conditions. Clinical validity of delayed CBC analysis depends on the clinical situation and required precision of the result.
AB - Introduction: This study assessed the comparability of complete blood count (CBC) parameters between capillary and venous samples, and extended previous research by examining the influence of different storage temperatures on CBC stability up to 7 days after sample collection. Methods: Venous and capillary blood samples were collected from 93 adult patients. Hemoglobin (Hb), hematocrit (Ht), mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), mean corpuscular hemoglobin concentration (MCHC), mean platelet volume (MPV), leukocytes, lymphocytes, basophils, eosinophils, erythrocytes, red cell distribution width (RDW), immature granulocytes (IG), immature reticulocyte fraction (IRF), monocytes, neutrophils, platelets, and reticulocytes were measured. Deming regression and mean relative differences between venous and capillary measurements were contrasted with desirable total allowable error (TEa). Stability was assessed in 20–27 venous blood samples stored at 4, 21–22, or 30°C, and analyzed at 0, 24, 48, 72, 96, 120, 144, and 168 h. Mean relative change with respect to baseline measurements was compared to the desirable TEa to determine acceptable stability. Results: Deming regression demonstrated strong linear correlations and acceptable variation between venous and capillary measurements. Erythrocytes, Hb, Ht, MCH, MCV, RDW, reticulocytes, and platelets showed acceptable stability for at least 96 h at 4°C. Mean relative change exceeded desirable TEa after 24 h at 30°C for all parameters, except erythrocytes, Hb, leukocytes, and MCH. Conclusion: Clinical laboratory specialists and clinicians should be aware of potential differences between venous and capillary measurements, and the influence of storage conditions. Clinical validity of delayed CBC analysis depends on the clinical situation and required precision of the result.
KW - analytical variation
KW - clinical laboratory techniques
KW - complete blood count
KW - hematology
KW - storage temperature
UR - http://www.scopus.com/inward/record.url?scp=85157989116&partnerID=8YFLogxK
U2 - 10.1111/ijlh.14080
DO - 10.1111/ijlh.14080
M3 - Article
C2 - 37128804
SN - 1751-5521
VL - 45
SP - 659
EP - 667
JO - International Journal of Laboratory Hematology
JF - International Journal of Laboratory Hematology
IS - 5
ER -