Abstract
Background: Liqoseal consists of a watertight layer of poly(ester)ether urethane and an adhesive layer containing polyethylene glycol-N-hydroxysuccinimide (PEG-NHS). It is designed to prevent cerebrospinal fluid (CSF) leakage after intradural surgery. This study assessed the safety and biodegradability of Liqoseal in a porcine craniotomy model.
Methods: In 32 pigs a craniotomy plus durotomy was performed. In 15 pigs Liqoseal was implanted, in 11 control pigs no sealant was implanted and in 6 control pigs a control dural sealant (Duraseal or Tachosil) was implanted. The safety of Liqoseal was evaluated by clinical, MRI and histological assessment. The degradation of Liqoseal was histologically estimated.
Results: Liqoseal, 2 mm thick before application, did not swell and significantly was at maximum mean thickness of 2.14 (±0.37) mm at one month. The foreign body reaction induced by Liqoseal, Duraseal and Tachosil were comparable. Liqoseal showed no adherence to the arachnoid layer and was completely resorbed between 6 and 12 months postoperatively. In one animal with Liqoseal, an epidural fluid collection containing CSF could not be excluded.
Conclusion: Liqoseal seems to be safe for intracranial use and is biodegradable. The safety and performance in humans needs to be further assessed in clinical trials.
| Original language | English |
|---|---|
| Pages (from-to) | 398-405 |
| Number of pages | 8 |
| Journal | Animal models and experimental medicine |
| Volume | 4 |
| Issue number | 4 |
| DOIs | |
| Publication status | Published - Dec 2021 |
Keywords
- cerebrospinal fluid leakage
- craniotomy
- dura
- in vivo model
- sealant