Abstract
BACKGROUND: Anastomotic connectors could be the key to less invasive bypass surgery, including endoscopic procedures, but equivalence to conventional suturing needs to be established. A novel distal coronary connector was tested pre-clinically for safety and efficacy in comparison to conventional suturing.
METHODS AND RESULTS: Left internal thoracic to left anterior descending coronary artery bypasses were constructed off-pump in 35 pigs (73+/-8 kg). An intraluminal metal connector (S2AS) was used in 21 and conventional suturing in 14 animals. S2AS anastomosis construction was easier achieved in one-fourth of the conventional construction time (3.7+/-0.7 versus 16.5+/-2.6 minutes; P<0.001). Acute patency tended to be better (P=0.15). All anastomoses were evaluated intraoperatively, and subgroups at 90 and at 180 days. Patency was 100%. An effective remodeling response was observed in all groups, resulting in unobstructed anastomoses with excellent hemodynamic performance (fractional flow reserve > or = 0.93 at 180 days). At 6 months, the noncompliant connector was covered with stabilized neointima that was thinner than found on the suture line (0.10+/-0.04 versus 0.31+/-0.13 mm; P=0.01). The connector induced less lumen loss (-0.6+/-6.5 versus 21.6+/-19%; P=0.03). The initial side-to-side configuration had remodeled to an end-to-side shape as intended.
CONCLUSIONS: In the porcine model, the connector rapidly and consistently produced high-quality anastomoses that fully met current standards on patency and function. Unconventional aspects like a noncompliant intraluminal ring and a side-to-side to end-to-side converted configuration did not interfere with favorable anastomosis remodeling. These findings shed a new light on the anatomical prerequisites for anastomosis patency.
| Original language | English |
|---|---|
| Pages (from-to) | I-390-I-395 |
| Number of pages | 6 |
| Journal | Circulation |
| Volume | 114 |
| Issue number | 1 suppl |
| DOIs | |
| Publication status | Published - 4 Jul 2006 |
Keywords
- Animals
- Coronary Angiography
- Coronary Artery Bypass, Off-Pump
- Corrosion Casting
- Equipment Design
- Female
- Follow-Up Studies
- Foreign-Body Reaction
- Graft Occlusion, Vascular
- Hemodynamics
- Implants, Experimental
- Internal Mammary-Coronary Artery Anastomosis
- Postoperative Complications
- Surgical Staplers
- Surgical Stapling
- Sus scrofa
- Suture Techniques
- Vascular Patency
- Wound Healing
Fingerprint
Dive into the research topics of 'S2 connector versus suture: distal coronary anastomosis remodeling, patency, and function in the pig'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver