TY - JOUR
T1 - Reliability and Agreement of 3D Anthropometric Measurements in Facial Palsy Patients Using a Low-Cost 4D Imaging System
AU - Harkel, Timen C Ten
AU - Vinayahalingam, Shankeeth
AU - Ingels, Koen J A O
AU - Berge, Stefaan J
AU - Maal, Thomas J J
AU - Speksnijder, Caroline M
N1 - Publisher Copyright:
© 2001-2011 IEEE.
Copyright:
Copyright 2021 Elsevier B.V., All rights reserved.
PY - 2020/8
Y1 - 2020/8
N2 - The reliability (precision) and agreement (accuracy) of anthropometric measurements based on manually placed 3D landmarks using the RealSense D415 were investigated in this paper. Thirty facial palsy patients, with their face in neutral (resting) position, were recorded simultaneously with the RealSense and a professional 3dMD imaging system. First the RealSense depth accuracy was determined. Subsequently, two observers placed 14 facial landmarks on the 3dMD and RealSense image, assessing the distance between landmark placement. The respective intra- and inter-rater Euclidean distance between the landmark placements was 0.84 mm (±0.58) and 1.00 mm (±0.70) for the 3dMD landmarks and 1.32 mm (±1.27) and 1.62 mm (±1.42) for the RealSense landmarks. From these landmarks 14 anthropometric measurements were derived. The intra- and inter-rater measurements had an overall reliability of 0.95 (0.87 - 0.98) and 0.93 (0.85 - 0.97) for the 3dMD measurements, and 0.83 (0.70 - 0.91) and 0.80 (0.64 - 0.89) for the RealSense measurements, respectively, expressed as the intra-class correlation coefficient. Determined by the Bland-Altman analysis, the agreement between the RealSense measurements and 3dMD measurements was on average -0.90 mm (-4.04 - 2.24) and -0.89 mm (-4.65 - 2.86) for intra- and inter-rater agreement, respectively. Based on the reported reliability and agreement of the RealSense measurements, the RealSense D415 can be considered as a viable option to perform objective 3D anthropomorphic measurements on the face in a neutral position, where a low-cost and portable camera is required.
AB - The reliability (precision) and agreement (accuracy) of anthropometric measurements based on manually placed 3D landmarks using the RealSense D415 were investigated in this paper. Thirty facial palsy patients, with their face in neutral (resting) position, were recorded simultaneously with the RealSense and a professional 3dMD imaging system. First the RealSense depth accuracy was determined. Subsequently, two observers placed 14 facial landmarks on the 3dMD and RealSense image, assessing the distance between landmark placement. The respective intra- and inter-rater Euclidean distance between the landmark placements was 0.84 mm (±0.58) and 1.00 mm (±0.70) for the 3dMD landmarks and 1.32 mm (±1.27) and 1.62 mm (±1.42) for the RealSense landmarks. From these landmarks 14 anthropometric measurements were derived. The intra- and inter-rater measurements had an overall reliability of 0.95 (0.87 - 0.98) and 0.93 (0.85 - 0.97) for the 3dMD measurements, and 0.83 (0.70 - 0.91) and 0.80 (0.64 - 0.89) for the RealSense measurements, respectively, expressed as the intra-class correlation coefficient. Determined by the Bland-Altman analysis, the agreement between the RealSense measurements and 3dMD measurements was on average -0.90 mm (-4.04 - 2.24) and -0.89 mm (-4.65 - 2.86) for intra- and inter-rater agreement, respectively. Based on the reported reliability and agreement of the RealSense measurements, the RealSense D415 can be considered as a viable option to perform objective 3D anthropomorphic measurements on the face in a neutral position, where a low-cost and portable camera is required.
KW - 3D anthropometric measurements
KW - 3dMD
KW - RealSense D415
KW - facial palsy
KW - intra-and inter-rater reliability and agreement
UR - http://www.scopus.com/inward/record.url?scp=85089205024&partnerID=8YFLogxK
U2 - 10.1109/TNSRE.2020.3007532
DO - 10.1109/TNSRE.2020.3007532
M3 - Article
C2 - 32746313
SN - 1534-4320
VL - 28
SP - 1817
EP - 1824
JO - IEEE Transactions on Neural Systems and Rehabilitation Engineering
JF - IEEE Transactions on Neural Systems and Rehabilitation Engineering
IS - 8
M1 - 9133557
ER -