Precision, accuracy and linearity of radiometer EML 105 whole blood metabolite biosensors

C Cobbaert, C Morales, M van Fessem, H Kemperman

    Research output: Contribution to journalArticleAcademicpeer-review

    Abstract

    The analytical performance of a new, whole blood glucose and lactate electrode system (EML 105 analyser. Radiometer Medical A/S. Copenhagen, Denmark) was evaluated. Between-day coefficients of variation were < or = 1.9% and < or = 3.1% for glucose and lactate, respectively. Recoveries of glucose were 100 +/- 10% using either aqueous or protein-based standards. Recoveries of lactate depended on the matrix, being underestimated in aqueous standards (approximately -10%) and 95-100% in standards containing 40 g/L albumin at lactate concentrations of 15 and 30 mmol/L. However, recoveries were high (up to 180%) at low lactate concentrations in protein-based standards. Carry-over, investigated according to National Clinical Chemistry Laboratory Standards EP10-T2, was negligible (alpha = 0.01). Glucose and lactate biosensors equipped with new membranes were linear up to 60 and 30 mmol/L, respectively. However, linearity fell upon daily use with increasing membrane lifetime. We conclude that the Radiometer metabolite biosensor results are reproducible and do not suffer from specimen-related carry-over. However, lactate recovery depends on the protein content and the lactate concentration.

    Original languageEnglish
    Pages (from-to)730-8
    Number of pages9
    JournalAnnals of Clinical Biochemistry
    Volume36 ( Pt 6)
    Publication statusPublished - Nov 1999

    Keywords

    • Biosensing Techniques
    • Blood Glucose
    • Humans
    • Lactic Acid
    • Radiometry
    • Sensitivity and Specificity

    Fingerprint

    Dive into the research topics of 'Precision, accuracy and linearity of radiometer EML 105 whole blood metabolite biosensors'. Together they form a unique fingerprint.

    Cite this