Practical considerations of diffusion-weighted MRS with ultra-strong diffusion gradients

Christopher W Davies-Jenkins*, André Döring, Fabrizio Fasano, Elena Kleban, Lars Mueller, C John Evans, Maryam Afzali, Derek K Jones, Itamar Ronen, Francesca Branzoli, Chantal M W Tax

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

3 Downloads (Pure)

Abstract

INTRODUCTION: Diffusion-weighted magnetic resonance spectroscopy (DW-MRS) offers improved cellular specificity to microstructure-compared to water-based methods alone-but spatial resolution and SNR is severely reduced and slow-diffusing metabolites necessitate higher b-values to accurately characterize their diffusion properties. Ultra-strong gradients allow access to higher b-values per-unit time, higher SNR for a given b-value, and shorter diffusion times, but introduce additional challenges such as eddy-current artefacts, gradient non-uniformity, and mechanical vibrations.

METHODS: In this work, we present initial DW-MRS data acquired on a 3T Siemens Connectom scanner equipped with ultra-strong (300 mT/m) gradients. We explore the practical issues associated with this manner of acquisition, the steps that may be taken to mitigate their impact on the data, and the potential benefits of ultra-strong gradients for DW-MRS. An in-house DW-PRESS sequence and data processing pipeline were developed to mitigate the impact of these confounds. The interaction of TE, b-value, and maximum gradient amplitude was investigated using simulations and pilot data, whereby maximum gradient amplitude was restricted. Furthermore, two DW-MRS voxels in grey and white matter were acquired using ultra-strong gradients and high b-values.

RESULTS: Simulations suggest T2-based SNR gains that are experimentally confirmed. Ultra-strong gradient acquisitions exhibit similar artefact profiles to those of lower gradient amplitude, suggesting adequate performance of artefact mitigation strategies. Gradient field non-uniformity influenced ADC estimates by up to 4% when left uncorrected. ADC and Kurtosis estimates for tNAA, tCho, and tCr align with previously published literature.

DISCUSSION: In conclusion, we successfully implemented acquisition and data processing strategies for ultra-strong gradient DW-MRS and results indicate that confounding effects of the strong gradient system can be ameliorated, while achieving shorter diffusion times and improved metabolite SNR.

Original languageEnglish
Article number1258408
Number of pages18
JournalFrontiers in Neuroscience
Volume17
DOIs
Publication statusPublished - 7 Dec 2023

Keywords

  • diffusion-weighted MRS
  • eddy currents
  • gradient non-uniformity
  • metabolites
  • ultra-strong gradients