Skip to main navigation Skip to search Skip to main content

Poly(L-lactide) bone plates and screws for internal fixation of mandibular swing osteotomies

  • J Tams
  • , F R Rozema
  • , R R Bos
  • , J L Roodenburg
  • , P G Nikkels
  • , A Vermey

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

This study evaluated bone healing after mandibular swing osteotomies fixed with biodegradable poly(L-lactide) (PLLA) bone plates in four patients. A step osteotomy treated with two PLLA bone plates (n=3), and a straight osteotomy treated with one PLLA bone plate (n=1) were performed. Bone healing was uneventful in all patients; only in the patient with the straight osteotomy was callus observed. After 5.5 years, radiologic changes at the site of implantation were observed in the remaining patient. Histologic examination revealed a nonspecific foreign-body reaction on highly crystalline PLLA remnants which were still present. It is concluded that the PLLA bone plates provided enough strength to enable undisturbed bone healing. The long-term degradation results are comparable with those of other studies on as-polymerized PLLA implants.

Original languageEnglish
Pages (from-to)20-4
Number of pages5
JournalInternational Journal of Oral and Maxillofacial Surgery
Volume25
Issue number1
Publication statusPublished - Feb 1996

Keywords

  • Biodegradation, Environmental
  • Bone Plates
  • Bone Screws
  • Bony Callus
  • Carcinoma, Squamous Cell
  • Crystallization
  • Equipment Design
  • Female
  • Follow-Up Studies
  • Foreign-Body Reaction
  • Humans
  • Internal Fixators
  • Male
  • Mandible
  • Middle Aged
  • Mouth Neoplasms
  • Oropharyngeal Neoplasms
  • Osteotomy
  • Polyesters
  • Wound Healing

Fingerprint

Dive into the research topics of 'Poly(L-lactide) bone plates and screws for internal fixation of mandibular swing osteotomies'. Together they form a unique fingerprint.