Near-infrared Spectroscopy to Indicate Selective Shunt Use During Carotid Endarterectomy

C.W.A. Pennekamp, R.V. Immink, H.M. den Ruijter, L.J. Kappelle, M.L. Bots, W.F. Buhre, F.L. Moll, G.J. de Borst

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

Objectives: This study assessed the value of cerebral near-infrared spectroscopy (NIRS) and transcranial Doppler (TCD) in relation to electroencephalography (EEG) changes for the detection of cerebral hypoperfusion necessitating shunt placement during carotid endarterectomy (CEA).

Methods: This was a prospective cohort study. Patients with a sufficient TCD window undergoing CEA from February 2009 to June 2011 were included. All patients were continuously monitored with NIRS and EEG. An intraluminal shunt was placed, selectively determined by predefined EEG changes in alpha, beta, theta, or delta activity. Relative changes in regional cerebral oxygen saturation (rSO(2)) in the frontal lobe and mean blood flow velocity (V-mean), 1 30 seconds before carotid cross-clamping versus 2 minutes after carotid cross-clamping were related to shunt placement. Receiver operating characteristic curve analysis was performed to determine the optimal thresholds. Diagnostic values were reported as positive and negative predictive value (PPV and NPV).

Results: Of a cohort of 151 patients, 17(11%) showed EEG changes requiring shunt placement. The rSO(2) and V-mean decreased more in the shunt group than in the non-shunt group (mean +/- standard error of the mean) 21 +/- 4% versus 7 +/- 5% and 76 +/- 6% versus 12 +/- 3%, respectively (p <.005), Receiver operating characteristic curve analysis revealed a threshold of 16% decrease in rSO(2) (PPV 76% and NPV 99%) and 48% decrease in V-mean (PPV 53% and NPV 99%) as the optimal cut-off value to detect cerebral ischemia during CEA under general anesthesia.

Conclusions: Compared with EEG, we found moderate PPV but high NPV for NIRS and TCD to detect cerebral ischemia during CEA under general anesthesia, meaning that both techniques independently may be suitable to exclude patients for unnecessary shunt use and to direct the use of selective shunting. However, the optimal thresholds for NIRS remain to be determined. (C) 2013 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.

Original languageEnglish
Pages (from-to)397-403
Number of pages7
JournalEuropean Journal of Vascular and Endovascular Surgery
Volume46
Issue number4
DOIs
Publication statusPublished - Oct 2013

Keywords

  • Carotid endarterectomy
  • Near-infrared spectroscopy
  • Selective shunt placement
  • Complications of carotid endarterectomy
  • RANDOMIZED CONTROLLED-TRIAL
  • GENERAL-ANESTHESIA
  • CEREBRAL OXIMETRY
  • TRANSCRANIAL DOPPLER
  • EEG CHANGES
  • ELECTROENCEPHALOGRAPHY
  • PREVENTION
  • OCCLUSION
  • PRESSURE
  • BENEFITS

Fingerprint

Dive into the research topics of 'Near-infrared Spectroscopy to Indicate Selective Shunt Use During Carotid Endarterectomy'. Together they form a unique fingerprint.

Cite this