MR-OCTAVIUS 4D with 1500 MR and 1600 MR arrays is suitable for plan QA in a 1.5 T MRI-linac

Viktoriia Gorobets, Wilfred de Vries, Nicole Brand, Thomas Foppen, André J.M. Wopereis, Simon Woodings*

*Corresponding author for this work

Research output: Contribution to journalReview articlepeer-review

Abstract

To ensure the accuracy of radiation delivery to patients in a 1.5 T MRI-linac, the implementation of quality assurance (QA) devices compatible with MR technology is essential. The OCTAVIUS 4D MR, made by PTW (Freiburg, Germany) is designed to ensure consistent and ideal alignment of its detectors with the direction of each beam segment. This study focuses on investigating the fundamental characteristics of the detector response for the OCTAVIUS Detector (OD) 1500 MR and OCTAVIUS 1600 MR when used in the MR-compatible OCTAVIUS 4D. Characteristics examined included short-term reproducibility, dose linearity, field size dependency, monitor unit (MU) rate dependency, dose-per-pulse dependency, and angular dependency. The evaluation of OD 1500 MR also involved measuring 25 clinical treatment plans across diverse target sizes and anatomical sites, including the liver/pancreas, rectum, prostate, lungs, and lymph nodes. One plan was measured with the standard setup and with a 5 cm left offset. The OD 1600 MR was not available for these measurements. The capability of the OD 1500 MR to identify potential errors was assessed by introducing a MU and positional shift within the software. The results demonstrated no significant differences in short-term reproducibility ( < 0.2 % ), dose linearity ( < 1 % ), field size dependency ( < 0.7 % for field sizes larger than 5 cm × 5 cm), MU rate dependency ( < 0.8 % ), dose-per-pulse dependency ( < 0.4 % ) and angular dependency (standard deviation < 0.5 % ). All tests of clinical plans were successfully completed. The OD 1500 MR demonstrated compatibility with the standard 95% pass rate when employing a global 3%/3 mm gamma criterion, and a 90% pass rate using a global 2%/2 mm gamma criterion. The detector demonstrated the capacity to measure treatment plans with a 5 cm left offset. With the standard parameters, the gamma test was sensitive to position errors but required an addition tests of mean/median dose or point dose in order to detect small dose difference.

Original languageEnglish
Article number17NT01
JournalPhysics in medicine and biology
Volume69
Issue number17
DOIs
Publication statusPublished - 7 Sept 2024

Keywords

  • MR-Linac
  • OCTAVIUS 4D Phantom MR
  • OCTAVIUS Detector 1500 MR
  • OCTAVIUS Detector 1600 MR
  • patient specific quality assurance

Fingerprint

Dive into the research topics of 'MR-OCTAVIUS 4D with 1500 MR and 1600 MR arrays is suitable for plan QA in a 1.5 T MRI-linac'. Together they form a unique fingerprint.

Cite this