Local SAR Assessment for Multitransmit Systems: A Study on the Peak Local SAR Value as a Function of Magnetic Field Strength

Alexander J. E. Raaijmakers, Bart R. Steensma

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

With increasing permanent magnetic field strength B 0, the frequency of the radiofrequency field B 1 increases. The concomitantly reduced wavelength results in interference patterns that cause signal inhomogeneities and for ultrahigh field strengths even signal voids. These inhomogeneities are addressed by multitransmit systems. These systems use multiple transmit elements where the phase and amplitude of each element can be controlled to steer the B 1‐field. However, in this way the electric field that originates from the B 1‐field also changes. This may pose a safety risk in terms of localized tissue heating. To avoid potential hotspots, the local specific absorption rate (SAR) distribution needs to be determined by simulations. All concepts and methods to process simulation results are presented in this article. In addition, we use the presented methodology to investigate the relationship between the peak local SAR value and the magnetic field strength B 0. For this purpose, we have simulated the birdcage body coil at 3 T and appropriately designed 16‐channel dipole‐loop transceiver arrays at 3, 7, 10.5, and 14 T. We demonstrate a linear increase of peak local SAR with B 0 and a quadratic increase of signal‐to‐noise‐ratio (SNR) with B 0.
Original languageEnglish
Pages (from-to)1-10
Number of pages10
JournaleMagRes
Volume8
Issue number1
DOIs
Publication statusPublished - 1 Jan 2019

Keywords

  • Dipole antennas
  • High-field imaging
  • MR imaging
  • Multitransmit
  • Sar
  • SAR
  • high-field imaging
  • dipole antennas
  • multitransmit

Fingerprint

Dive into the research topics of 'Local SAR Assessment for Multitransmit Systems: A Study on the Peak Local SAR Value as a Function of Magnetic Field Strength'. Together they form a unique fingerprint.

Cite this