T1 - Lack of Prognostic Value of T-Wave Alternans for Implantable Cardioverter-Defibrillator Benefit in Primary Prevention
AU - Pelli, Ari
AU - Kenttä, Tuomas V.
AU - Junttila, M. Juhani
AU - Huber, Cynthia
AU - Schlögl, Simon
AU - Zabel, Markus
AU - Malik, Marek
AU - Willems, Rik
AU - Vos, Marc A.
AU - Harden, Markus
AU - Friede, Tim
AU - Sticherling, Christian
AU - Huikuri, Heikki V.
N1 - Publisher Copyright:
© 2024 The Authors.
PY - 2024/6/4
Y1 - 2024/6/4
N2 - BACKGROUND: New methods to identify patients who benefit from a primary prophylactic implantable cardioverter-defibrillator (ICD) are needed. T-wave alternans (TWA) has been shown to associate with arrhythmogenesis of the heart and sudden cardiac death. We hypothesized that TWA might be associated with benefit from ICD implantation in primary prevention. METHODS AND RESULTS: In the EU-CERT-ICD (European Comparative Effectiveness Research to Assess the Use of Primary Prophylactic Implantable Cardioverter-Defibrillators) study, we prospectively enrolled 2327 candidates for primary prophylactic ICD. A 24-hour Holter monitor reading was taken from all recruited patients at enrollment. TWA was assessed from Holter monitoring using the modified moving average method. Study outcomes were all-cause death, appropriate shock, and survival benefit. TWA was assessed both as a contiguous variable and as a dichotomized variable with cutoff points <47 μV and <60 μV. The final cohort included 1734 valid T-wave alternans samples, 1211 patients with ICD, and 523 control patients with conservative treatment, with a mean follow-up time of 2.3 years. TWA ≥60 μV was a predicter for a higher all-cause death in patients with an ICD on the basis of a univariate Cox regression model (hazard ratio, 1.484 [95% CI, 1.024–2.151]; P=0.0374; concordance statistic, 0.51). In multivariable models, TWA was not prognostic of death or appropriate shocks in patients with an ICD. In addition, TWA was not prognostic of death in control patients. In a propensity score–adjusted Cox regression model, TWA was not a predictor of ICD benefit. CONCLUSIONS: T-wave alternans is poorly prognostic in patients with a primary prophylactic ICD. Although it may be prognostic of life-threatening arrhythmias and sudden cardiac death in several patient populations, it does not seem to be useful in assessing benefit from ICD therapy in primary prevention among patients with an ejection fraction of ≤35%.
AB - BACKGROUND: New methods to identify patients who benefit from a primary prophylactic implantable cardioverter-defibrillator (ICD) are needed. T-wave alternans (TWA) has been shown to associate with arrhythmogenesis of the heart and sudden cardiac death. We hypothesized that TWA might be associated with benefit from ICD implantation in primary prevention. METHODS AND RESULTS: In the EU-CERT-ICD (European Comparative Effectiveness Research to Assess the Use of Primary Prophylactic Implantable Cardioverter-Defibrillators) study, we prospectively enrolled 2327 candidates for primary prophylactic ICD. A 24-hour Holter monitor reading was taken from all recruited patients at enrollment. TWA was assessed from Holter monitoring using the modified moving average method. Study outcomes were all-cause death, appropriate shock, and survival benefit. TWA was assessed both as a contiguous variable and as a dichotomized variable with cutoff points <47 μV and <60 μV. The final cohort included 1734 valid T-wave alternans samples, 1211 patients with ICD, and 523 control patients with conservative treatment, with a mean follow-up time of 2.3 years. TWA ≥60 μV was a predicter for a higher all-cause death in patients with an ICD on the basis of a univariate Cox regression model (hazard ratio, 1.484 [95% CI, 1.024–2.151]; P=0.0374; concordance statistic, 0.51). In multivariable models, TWA was not prognostic of death or appropriate shocks in patients with an ICD. In addition, TWA was not prognostic of death in control patients. In a propensity score–adjusted Cox regression model, TWA was not a predictor of ICD benefit. CONCLUSIONS: T-wave alternans is poorly prognostic in patients with a primary prophylactic ICD. Although it may be prognostic of life-threatening arrhythmias and sudden cardiac death in several patient populations, it does not seem to be useful in assessing benefit from ICD therapy in primary prevention among patients with an ejection fraction of ≤35%.
KW - appropriate shock
KW - death
KW - heart failure
KW - implantable cardioverter-defibrillator
KW - T-wave alternans
UR - http://www.scopus.com/inward/record.url?scp=85195018331&partnerID=8YFLogxK
U2 - 10.1161/JAHA.123.032465
DO - 10.1161/JAHA.123.032465
M3 - Article
C2 - 38804218
AN - SCOPUS:85195018331
SN - 2047-9980
VL - 13
JO - Journal of the American Heart Association
JF - Journal of the American Heart Association
IS - 11
M1 - e032465
ER -