Skip to main navigation Skip to search Skip to main content

Interobserver reproducibility of qualitative and quantitative analysis of intravascular ultrasound images before and after peripheral balloon angioplasty

  • Aad Van Der Lugt
  • , Elma J. Gussenhoven*
  • , Gerard Pasterkamp
  • , Nicolaas Bom
  • , David J. Posthuma
  • , Theo Stijnen
  • *Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

11 Citations (Scopus)

Abstract

In this study, interobserver agreement on intravascular ultrasound data obtained before and after balloon angioplasty (PTA) of the superficial femoral artery was assessed. Two observers analyzed intravascular ultrasound cross-sections from 38 patients. Interobserver agreement was good for soft and hard lesions, dissection and vascular damage (kappa values 0.61, 0.67, 0.69 and 0.66, respectively); moderate for eccentric lesions (kappa 0.45); fair for media ruptures (kappa 0.25); and poor for lipid deposits and plaque raptures (kappa 0.0 and 0,04, respectively). Differences for the are of normal wall and hard lesion were not significant, but were for dissection. There were no significant interobserver differences between area measurements. The coefficient of variation for free lumen area and media- bounded area before PTA was 17.2% and 10.5% and after PTA 11.2% and 9.2%, respectively. This study identified the intravascular ultrasound parameters that are reproducible.

Original languageEnglish
Pages (from-to)399-404
Number of pages6
JournalUltrasound in Medicine and Biology
Volume22
Issue number4
DOIs
Publication statusPublished - 1 Jan 1996

Keywords

  • Balloon angioplasty
  • Intravascular ultrasound
  • Peripheral arteries
  • Reproducibility