T1 - In vivo and in vitro relationship between ionized magnesium and ionized calcium
AU - Groenestege, Wouter M.Tiel
AU - Stokwielder, Ron H.
AU - Soels, Leosa R.
AU - Sikma, Maaike A.
AU - Hutten, Tim J.A.
N1 - Publisher Copyright:
© 2024 The Author(s)
PY - 2024/12
Y1 - 2024/12
N2 - Objectives: The objective of this study was to determine the in vivo correlation of ionized magnesium (iMg) with ionized calcium (iCa), total calcium, albumin and pH. In addition, the analytical interference of iCa on iMg measurement on the Stat Profile Prime Plus (Nova Biomedical) and vice versa was defined. Methods: In vivo correlation of iCa, iMg and pH was studied in 238 paired blood gas samples of 109 different patients admitted to the intensive care unit. Albumin and total magnesium (tMg) were measured in heparinized plasma samples. Measurement of iMg was performed with the ion selective magnesium electrode (ISE) of the Stat Profile Prime Plus (Nova Biomedical) and iCa and pH were measured with a Rapid Point 500 blood gas analyzer (Siemens). Albumin, total calcium and total magnesium were analyzed with a Siemens Atellica CH. Analytical interference of iCa with iMg and vice versa was investigated using unbuffered saline solutions. Results: In the studied patient population, no significant correlations were observed between iMg and iCa, albumin, and pH. An inverse relationship was observed between iCa and Mg-ISE. For every 0.1 mmol/L change in iCa concentration, the iMg concentration deviated by 0.01 mmol/L at an iMg concentration of 0.5 mmol/L and by 0.013 mmol/L at an iMg concentration of 1.0 mmol/L. The measurement of iCa was not affected by iMg. Conclusions: In vivo, no correlation was observed between iMg with iCa, albumin and pH. Interference of iCa on iMg measurement was noted, with a maximum deviation of ±0.02 mmol/L iMg across the reference range of iCa (1.15–1.32 mmol/L). Additionally, the iCa measurement was not affected by the iMg concentration.
AB - Objectives: The objective of this study was to determine the in vivo correlation of ionized magnesium (iMg) with ionized calcium (iCa), total calcium, albumin and pH. In addition, the analytical interference of iCa on iMg measurement on the Stat Profile Prime Plus (Nova Biomedical) and vice versa was defined. Methods: In vivo correlation of iCa, iMg and pH was studied in 238 paired blood gas samples of 109 different patients admitted to the intensive care unit. Albumin and total magnesium (tMg) were measured in heparinized plasma samples. Measurement of iMg was performed with the ion selective magnesium electrode (ISE) of the Stat Profile Prime Plus (Nova Biomedical) and iCa and pH were measured with a Rapid Point 500 blood gas analyzer (Siemens). Albumin, total calcium and total magnesium were analyzed with a Siemens Atellica CH. Analytical interference of iCa with iMg and vice versa was investigated using unbuffered saline solutions. Results: In the studied patient population, no significant correlations were observed between iMg and iCa, albumin, and pH. An inverse relationship was observed between iCa and Mg-ISE. For every 0.1 mmol/L change in iCa concentration, the iMg concentration deviated by 0.01 mmol/L at an iMg concentration of 0.5 mmol/L and by 0.013 mmol/L at an iMg concentration of 1.0 mmol/L. The measurement of iCa was not affected by iMg. Conclusions: In vivo, no correlation was observed between iMg with iCa, albumin and pH. Interference of iCa on iMg measurement was noted, with a maximum deviation of ±0.02 mmol/L iMg across the reference range of iCa (1.15–1.32 mmol/L). Additionally, the iCa measurement was not affected by the iMg concentration.
KW - Albumin
KW - Calcium
KW - Interference
KW - Ion selective electrode
KW - Ionized calcium
KW - Ionized magnesium
KW - pH
UR - http://www.scopus.com/inward/record.url?scp=85203197753&partnerID=8YFLogxK
U2 - 10.1016/j.clinbiochem.2024.110815
DO - 10.1016/j.clinbiochem.2024.110815
M3 - Article
AN - SCOPUS:85203197753
SN - 0009-9120
VL - 133-134
SP - 1
EP - 4
JO - Clinical biochemistry
JF - Clinical biochemistry
M1 - 110815
ER -