Abstract
Purpose: Recent public health safety issues involving medical devices have led to a growing demand to improve the current passive-reactive postmarket surveillance (PMS) system. Various European Union (EU) national competent authorities have started to focus on strengthening the postmarket risk evaluation. As a consequence, the new EU medical device regulation was published; it includes the concept of a PMS Plan. Methods: This publication reviewed Annex III Technical Documentation on PMS and Annex XIV Part B: Postmarket clinical follow-up from the new Regulation (EU) 2017/745 of the European Parliament and of the Council on medical devices. Results: The results of the PMS activities will be described in the PMS plan and will be used to update other related documents. A modular approach to structure the contents of the PMS plan will help to consistently update other PMS information. It is our suggestion that the PMS plan should consist of a PMS plan Core and a PMS plan Supplement. The PMS plan Core document will describe the PMS system, and the PMS plan Supplement will outline the specific activities performed by the manufacturer for a particular medical device. Conclusions: The PMS plan may serve as a thorough tool for the benefit-risk evaluation of medical devices. If properly developed and implemented, it will function as a key player in the establishment of a new framework for proactive safety evaluation of medical devices.
Original language | English |
---|---|
Pages (from-to) | 1155-1165 |
Number of pages | 11 |
Journal | Pharmacoepidemiology and Drug Safety |
Volume | 28 |
Issue number | 9 |
DOIs | |
Publication status | Published - 1 Sept 2019 |
Keywords
- medical devices
- pharmacoepidemiology
- postmarket surveillance plan
- risk management
- safety evaluation