Abstract
Background and purpose The latency of a multileaf collimator (MLC) tracking system used to overcome respiratory motion causes misalignment of the treatment beam with respect to the gross tumour volume, which may result in reduced target coverage. This study investigates the magnitude of this effect. Material and methods Simulated superior–inferior breathing motion was used to construct histograms of isocentre offset with respect to the gross tumour volume (GTV) for a variety of tracking latencies. Dose distributions for conformal volumetric modulated arc therapy (VMAT) arcs were then calculated at a range of offsets and summed according to these displacement histograms. The results were verified by delivering the plans to a Delta4 phantom on a motion platform. Results In the absence of an internal target margin, a tracking latency of 150 ms reduces the GTV D95% by approximately 2%. With a margin of 2 mm, the same drop in dose occurs for a tracking latency of 450 ms. Lung V13Gy is unaffected by a range of latencies. These results are supported by the phantom measurements. Conclusions Assuming that internal motion can be modelled by a rigid translation of the patient, MLC tracking of conformal VMAT can be effectively accomplished in the absence of an internal target margin for substantial breathing motion (4 s period and 20 mm peak–peak amplitude) so long as the system latency is less than 150 ms.
Original language | English |
---|---|
Pages (from-to) | 491-495 |
Number of pages | 5 |
Journal | Radiotherapy and Oncology |
Volume | 117 |
Issue number | 3 |
DOIs | |
Publication status | Published - 1 Dec 2015 |
Externally published | Yes |
Keywords
- Latency
- Lung
- MLC tracking
- Motion prediction
- VMAT