TY - JOUR
T1 - Direct measurement of ion chamber correction factors, k Q and k B, in a 7 MV MRI-linac
AU - de Prez, Leon
AU - Woodings, Simon
AU - de Pooter, Jacco
AU - van Asselen, Bram
AU - Wolthaus, Jochem
AU - Jansen, Bartel
AU - Raaymakers, Bas
PY - 2019/5/21
Y1 - 2019/5/21
N2 - The output of MRI-integrated photon therapy (MRgXT) devices is measured in terms of absorbed dose to water, D w. Traditionally this is done with reference type ion chambers calibrated in a beam quality Q 0 without magnetic field. To correct the ion chamber response for the application in the magnetic field, a factor needs to be applied that corrects for both beam quality Q and the presence of the magnetic field B, k Q,B. This can be expressed as the product of k Q, without magnetic field, and ion chamber magnetic field correction, k B. k B depends on the magnetic field strength and its direction, the direction of the beam and the orientation and type of the ion chamber. In this study, for the first time, both k Q and k B were measured directly for six waterproof ion chambers (3 × PTW 30013 and 3 × IBA FC65-G) in a pre-clinical 7 MV MRI-linac at 0 T and at 1.5 T. Measurements were done with the only available primary standard built for this purpose, a water calorimeter. Resulting k Q factors for PTW and IBA chambers were 0.985(5) and 0.990(4), respectively. k B factors were measured with the chambers in antiparallel direction to the magnetic field (|| 180°), and perpendicular direction (⊥ -90°). k B|| and k B⊥ for the PTW chambers were 0.985(6) and 0.963(4), respectively and for IBA chambers 0.995(4) and 0.956(4). Agreement with the available literature values was shown, partly caused by the relatively large standard deviation (SD) in those values. The values in this study are currently the only available measured values for k Q and k B in an MRI-linac that are directly linked to the international traceability framework for the quantity absorbed dose to water, D w.
AB - The output of MRI-integrated photon therapy (MRgXT) devices is measured in terms of absorbed dose to water, D w. Traditionally this is done with reference type ion chambers calibrated in a beam quality Q 0 without magnetic field. To correct the ion chamber response for the application in the magnetic field, a factor needs to be applied that corrects for both beam quality Q and the presence of the magnetic field B, k Q,B. This can be expressed as the product of k Q, without magnetic field, and ion chamber magnetic field correction, k B. k B depends on the magnetic field strength and its direction, the direction of the beam and the orientation and type of the ion chamber. In this study, for the first time, both k Q and k B were measured directly for six waterproof ion chambers (3 × PTW 30013 and 3 × IBA FC65-G) in a pre-clinical 7 MV MRI-linac at 0 T and at 1.5 T. Measurements were done with the only available primary standard built for this purpose, a water calorimeter. Resulting k Q factors for PTW and IBA chambers were 0.985(5) and 0.990(4), respectively. k B factors were measured with the chambers in antiparallel direction to the magnetic field (|| 180°), and perpendicular direction (⊥ -90°). k B|| and k B⊥ for the PTW chambers were 0.985(6) and 0.963(4), respectively and for IBA chambers 0.995(4) and 0.956(4). Agreement with the available literature values was shown, partly caused by the relatively large standard deviation (SD) in those values. The values in this study are currently the only available measured values for k Q and k B in an MRI-linac that are directly linked to the international traceability framework for the quantity absorbed dose to water, D w.
KW - MRgRT
KW - absorbed dose
KW - kQ
KW - magnetic field
KW - kB
KW - k
UR - http://www.scopus.com/inward/record.url?scp=85066457375&partnerID=8YFLogxK
U2 - 10.1088/1361-6560/ab1511
DO - 10.1088/1361-6560/ab1511
M3 - Article
C2 - 30933939
SN - 0031-9155
VL - 64
JO - Physics in Medicine and Biology
JF - Physics in Medicine and Biology
IS - 10
M1 - 105025
ER -