T1 - Comparison of Two Generations of Thoracic Aortic Stent Grafts and Their Impact on Aortic Stiffness in an Ex Vivo Porcine Model
AU - Mandigers, Tim J.
AU - Conti, Michele
AU - Allievi, Sara
AU - Dedola, Francesca
AU - Bissacco, Daniele
AU - Bianchi, Daniele
AU - Marconi, Stefania
AU - Domanin, Maurizio
AU - Van Herwaarden, Joost A.
AU - Auricchio, Ferdinando
AU - Trimarchi, Santi
N1 - Publisher Copyright:
© 2023 The Author(s)
PY - 2023/1
Y1 - 2023/1
N2 - Objective: Little is known about the cardiovascular changes after TEVAR and regarding the impact on aortic stiffness for different stent graft generations specifically, following changes in device design. The present study evaluated the stent graft induced aortic stiffening of two generations of the Valiant thoracic aortic stent graft. Methods: This was an ex vivo porcine investigation using an experimental mock circulatory loop. Thoracic aortas of young healthy pigs were harvested and connected to the mock circulatory loop. At a 60 bpm heart rate and stable mean arterial pressure, baseline aortic characteristics were obtained. Pulse wave velocity (PWV) was calculated before and after stent graft deployment. Paired and independent sample t tests or their non-parametric alternatives were performed to test for differences where appropriate. Results: Twenty porcine thoracic aortas were divided into two equal subgroups, in which a Valiant Captivia or a Valiant Navion stent graft was deployed. Both stent grafts were similar in diameter and length. Baseline aortic characteristics did not differ between the subgroups. Mean arterial pressure values did not change after either stent graft, while pulse pressures increased statistically significantly after Captivia (mean 44 ± 10 mmHg to 51 ± 13 mmHg, p = .002) but not after Navion. Mean baseline PWV increased after both Captivia (4.4 ± 0.6 m/s to 4.8 ± 0.7 m/s, p = .007) and Navion (4.6 ± 0.7 m/s to 4.9 ± 0.7 m/s, p = .002). There was no statistically significant difference in the mean percentage increase in PWV for either subgroup (8 ± 4% vs. 6 ± 4%, p = .25). Conclusion: These experimental findings showed no statistically significant difference in the percentage increase of aortic PWV after either stent graft generation and confirm that TEVAR increases aortic PWV. As a surrogate for aortic stiffness, this calls for further improvements in future thoracic aortic stent graft designs regarding device compliance.
AB - Objective: Little is known about the cardiovascular changes after TEVAR and regarding the impact on aortic stiffness for different stent graft generations specifically, following changes in device design. The present study evaluated the stent graft induced aortic stiffening of two generations of the Valiant thoracic aortic stent graft. Methods: This was an ex vivo porcine investigation using an experimental mock circulatory loop. Thoracic aortas of young healthy pigs were harvested and connected to the mock circulatory loop. At a 60 bpm heart rate and stable mean arterial pressure, baseline aortic characteristics were obtained. Pulse wave velocity (PWV) was calculated before and after stent graft deployment. Paired and independent sample t tests or their non-parametric alternatives were performed to test for differences where appropriate. Results: Twenty porcine thoracic aortas were divided into two equal subgroups, in which a Valiant Captivia or a Valiant Navion stent graft was deployed. Both stent grafts were similar in diameter and length. Baseline aortic characteristics did not differ between the subgroups. Mean arterial pressure values did not change after either stent graft, while pulse pressures increased statistically significantly after Captivia (mean 44 ± 10 mmHg to 51 ± 13 mmHg, p = .002) but not after Navion. Mean baseline PWV increased after both Captivia (4.4 ± 0.6 m/s to 4.8 ± 0.7 m/s, p = .007) and Navion (4.6 ± 0.7 m/s to 4.9 ± 0.7 m/s, p = .002). There was no statistically significant difference in the mean percentage increase in PWV for either subgroup (8 ± 4% vs. 6 ± 4%, p = .25). Conclusion: These experimental findings showed no statistically significant difference in the percentage increase of aortic PWV after either stent graft generation and confirm that TEVAR increases aortic PWV. As a surrogate for aortic stiffness, this calls for further improvements in future thoracic aortic stent graft designs regarding device compliance.
KW - Aortic stiffness
KW - Captivia
KW - Experimental investigation
KW - Navion
KW - Pulse wave velocity
KW - Valiant
UR - http://www.scopus.com/inward/record.url?scp=85159066720&partnerID=8YFLogxK
U2 - 10.1016/j.ejvsvf.2023.04.001
DO - 10.1016/j.ejvsvf.2023.04.001
M3 - Article
AN - SCOPUS:85159066720
VL - 59
SP - 8
EP - 14
JO - EJVES Vascular Forum
JF - EJVES Vascular Forum
ER -