T1 - Comparison of the GeneFinderTM COVID-19 Plus RealAmp Kit on the sample-to-result Platform ELITe InGenius to the national reference method
T2 - An added value of N gene target detection?
AU - Ong, David S Y
AU - Claas, Eric C J
AU - Breijer, Simone
AU - Vaessen, Norbert
N1 - Publisher Copyright:
© 2020 Elsevier B.V.
PY - 2020/11
Y1 - 2020/11
N2 - BACKGROUND: Due to the emergence of the coronavirus disease 2019 (COVID-19) pandemic there is an urgent need for rapid and accurate testing on the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2).OBJECTIVES: The aim of this study was to assess the diagnostic performance of the GeneFinderTMCOVID-19 Plus RealAmp Kit on the ELITe InGenius sample-to-result platform, which is a commercial nucleic acid amplification test (NAT) targeting genes of SARS-CoV-2.STUDY DESIGN: Patients were eligible between March 18 and May 27, 2020, when they had respiratory symptoms that were suspected for COVID-19. The InGenius platform was compared to routine in-house NAT that was validated according to the national reference.RESULTS: Of 128 randomly selected patients, 58 (45 %) tested positive and 55 (43 %) tested negative in both platforms. Sensitivity of the InGenius platform was 100 % (95 % confidence interval 94-100). In the remaining 15 (12 %) cases E and RdRp genes were not detected in both platforms but the nucleoprotein (N) gene was tested positive by the InGenius platform. All solitary N gene positive cases were confirmed by a N-gene specific in-house validated NAT, and most of these patients could also be considered positive based on other recently available COVID-19 positive respiratory samples or highly suspected radiological findings.CONCLUSION: The InGenius platform for SARS-CoV-2 detection has excellent sensitivity, is easy to use and provides fast results. The inclusion of the N gene as a third gene target may further increase sensitivity for the diagnosis of COVID-19 in comparison to the national reference method.
AB - BACKGROUND: Due to the emergence of the coronavirus disease 2019 (COVID-19) pandemic there is an urgent need for rapid and accurate testing on the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2).OBJECTIVES: The aim of this study was to assess the diagnostic performance of the GeneFinderTMCOVID-19 Plus RealAmp Kit on the ELITe InGenius sample-to-result platform, which is a commercial nucleic acid amplification test (NAT) targeting genes of SARS-CoV-2.STUDY DESIGN: Patients were eligible between March 18 and May 27, 2020, when they had respiratory symptoms that were suspected for COVID-19. The InGenius platform was compared to routine in-house NAT that was validated according to the national reference.RESULTS: Of 128 randomly selected patients, 58 (45 %) tested positive and 55 (43 %) tested negative in both platforms. Sensitivity of the InGenius platform was 100 % (95 % confidence interval 94-100). In the remaining 15 (12 %) cases E and RdRp genes were not detected in both platforms but the nucleoprotein (N) gene was tested positive by the InGenius platform. All solitary N gene positive cases were confirmed by a N-gene specific in-house validated NAT, and most of these patients could also be considered positive based on other recently available COVID-19 positive respiratory samples or highly suspected radiological findings.CONCLUSION: The InGenius platform for SARS-CoV-2 detection has excellent sensitivity, is easy to use and provides fast results. The inclusion of the N gene as a third gene target may further increase sensitivity for the diagnosis of COVID-19 in comparison to the national reference method.
KW - COVID-19 Testing/methods
KW - COVID-19/diagnosis
KW - Humans
KW - Molecular Diagnostic Techniques
KW - Nucleic Acid Amplification Techniques/methods
KW - Reproducibility of Results
KW - SARS-CoV-2/genetics
KW - Sensitivity and Specificity