TY - JOUR
T1 - Comparing clinical performance of current implantable cardioverter-defibrillator implantation recommendations in arrhythmogenic right ventricular cardiomyopathy
AU - Bosman, Laurens P
AU - Nielsen Gerlach, Claire L
AU - Cadrin-Tourigny, Julia
AU - Orgeron, Gabriela
AU - Tichnell, Crystal
AU - Murray, Brittney
AU - Bourfiss, Mimount
AU - van der Heijden, Jeroen F
AU - Yap, Sing-Chien
AU - Zeppenfeld, Katja
AU - van den Berg, Maarten P
AU - Wilde, Arthur A M
AU - Asselbergs, Folkert W
AU - Tandri, Hariskrishna
AU - Calkins, Hugh
AU - van Tintelen, J Peter
AU - James, Cynthia A
AU - Te Riele, Anneline S J M
N1 - Publisher Copyright:
© 2021 The Author(s) 2021. Published by Oxford University Press on behalf of the European Society of Cardiology.
PY - 2022/2
Y1 - 2022/2
N2 - AIMS: Arrhythmogenic right ventricular cardiomyopathy (ARVC) patients have an increased risk of ventricular arrhythmias (VA). Four implantable cardioverter-defibrillator (ICD) recommendation algorithms are available The International Task Force Consensus ('ITFC'), an ITFC modification by Orgeron et al. ('mITFC'), the AHA/HRS/ACC guideline for VA management ('AHA'), and the HRS expert consensus statement ('HRS'). This study aims to validate and compare the performance of these algorithms in ARVC.METHODS AND RESULTS: We classified 617 definite ARVC patients (38.5 ± 15.1 years, 52.4% male, 39.2% prior sustained VA) according to four algorithms. Clinical performance was evaluated by sensitivity, specificity, ROC-analysis, and decision curve analysis for any sustained VA and for fast VA (>250 b.p.m.). During 6.4 [2.8-11.5] years follow-up, 282 (45.7%) patients experienced any sustained VA, and 63 (10.2%) fast VA. For any sustained VA, ITFC and mITFC provide higher sensitivity than AHA and HRS (94.0-97.8% vs. 76.7-83.5%), but lower specificity (15.9-32.0% vs. 42.7%-60.1%). Similarly, for fast VA, ITFC and mITFC provide higher sensitivity than AHA and HRS (95.2-97.1% vs. 76.7-78.4%) but lower specificity (42.7-43.1 vs. 76.7-78.4%). Decision curve analysis showed ITFC and mITFC to be superior for a 5-year sustained VA risk ICD indication threshold between 5-25% or 2-9% for fast VA.CONCLUSION: The ITFC and mITFC provide the highest protection rates, whereas AHA and HRS decrease unnecessary ICD placements. ITFC or mITFC should be used if we consider the 5-year threshold for ICD indication to lie within 5-25% for sustained VA or 2-9% for fast VA. These data will inform decision-making for ICD placement in ARVC.
AB - AIMS: Arrhythmogenic right ventricular cardiomyopathy (ARVC) patients have an increased risk of ventricular arrhythmias (VA). Four implantable cardioverter-defibrillator (ICD) recommendation algorithms are available The International Task Force Consensus ('ITFC'), an ITFC modification by Orgeron et al. ('mITFC'), the AHA/HRS/ACC guideline for VA management ('AHA'), and the HRS expert consensus statement ('HRS'). This study aims to validate and compare the performance of these algorithms in ARVC.METHODS AND RESULTS: We classified 617 definite ARVC patients (38.5 ± 15.1 years, 52.4% male, 39.2% prior sustained VA) according to four algorithms. Clinical performance was evaluated by sensitivity, specificity, ROC-analysis, and decision curve analysis for any sustained VA and for fast VA (>250 b.p.m.). During 6.4 [2.8-11.5] years follow-up, 282 (45.7%) patients experienced any sustained VA, and 63 (10.2%) fast VA. For any sustained VA, ITFC and mITFC provide higher sensitivity than AHA and HRS (94.0-97.8% vs. 76.7-83.5%), but lower specificity (15.9-32.0% vs. 42.7%-60.1%). Similarly, for fast VA, ITFC and mITFC provide higher sensitivity than AHA and HRS (95.2-97.1% vs. 76.7-78.4%) but lower specificity (42.7-43.1 vs. 76.7-78.4%). Decision curve analysis showed ITFC and mITFC to be superior for a 5-year sustained VA risk ICD indication threshold between 5-25% or 2-9% for fast VA.CONCLUSION: The ITFC and mITFC provide the highest protection rates, whereas AHA and HRS decrease unnecessary ICD placements. ITFC or mITFC should be used if we consider the 5-year threshold for ICD indication to lie within 5-25% for sustained VA or 2-9% for fast VA. These data will inform decision-making for ICD placement in ARVC.
KW - Arrhythmogenic right ventricular cardiomyopathy
KW - Implantable cardioverter-defibrillator
KW - Prognosis
KW - Risk stratification
KW - Ventricular arrhythmias
UR - https://www.scopus.com/pages/publications/85125368890
U2 - 10.1093/europace/euab162
DO - 10.1093/europace/euab162
M3 - Article
C2 - 34468736
SN - 1099-5129
VL - 24
SP - 296
EP - 305
JO - Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology
JF - Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology
IS - 2
ER -