Abstract
The FlexitSystem implant is a novel implant used in open wedge high tibial osteotomy. A clinical safety study was performed. Retrospectively 50 patients were analyzed who were treated with an open wedge high tibial osteotomy and the new FlexitSystem implant, with a minimal follow-up of one year. Complication rate, radiographic outcomes and implant removal were investigated. One patient underwent a revision surgery because of loss of correction and non-union. The complication rate was 10.0%. No other radiographic complications (screw breakage, implant failure) were found. In 24 patients (48%) the FlexitSystem implant was removed at a mean follow-up of 12.6 months (range 2.6 till 24.0 months). The mean reason was irritation of the implant. The FlexitSystem implant is a clinical safe and stable implant for an open wedge high tibial osteotomy, with a low complication rate. The rate of implant irritation requiring removal remained high.
Original language | English |
---|---|
Pages (from-to) | 381-386 |
Number of pages | 6 |
Journal | Acta Orthopaedica Belgica |
Volume | 85 |
Issue number | 3 |
Publication status | Published - Sept 2019 |
Keywords
- Adult
- Aged
- Bone Plates
- Bone Screws
- Female
- Humans
- Male
- Middle Aged
- Osteoarthritis, Knee/surgery
- Osteotomy/adverse effects
- Prostheses and Implants
- Prosthesis Implantation/methods
- Retrospective Studies
- Tibia/surgery