Challenges in Removing the Essure® Device

Daniëlle M van Gastel, Liselotte W Maassen, Carolien A M Koks, Sebastiaan Veersema, Marlies Y Bongers

Research output: Contribution to journalArticleAcademicpeer-review

9 Downloads (Pure)

Abstract

We present a case about Essure® removal surgery in which the third markers of the device have torn off. The woman needed a second surgery for complete removal of the devices. Fluoroscopy during surgery is a good method to visualize the lost fragments. With fluoroscopy, a hysterectomy is not needed for complete removal. It is important to understand the structure of the device and to be aware of the four radiopaque markers during surgery and their removal.

Original languageEnglish
Article number8823606
Pages (from-to)1-4
JournalCase Reports in Obstetrics and Gynecology
Volume2020
DOIs
Publication statusPublished - 2020

Fingerprint

Dive into the research topics of 'Challenges in Removing the Essure® Device'. Together they form a unique fingerprint.

Cite this