Abstract
Little is known about the clinical application of highly cross-linked polyethylene (HXLPE) blended with vitamin E. This study evaluates an uncemented iso-elastic monoblock cup with vitamin E blended HXLPE. 112 patients were followed up for 2. years. 95.5% completed the follow-up. The mean VAS score for patient satisfaction was 8.8 and the mean Harris Hip Score was 94.2. In 7 cases initial gaps behind the cup were observed, which disappeared completely during follow-up in 6 cases. The mean femoral head penetration rate was 0.055. mm/year. No adverse reactions or abnormal mechanical behavior was observed with the short term use of vitamin E blended HXLPE. This study shows the promising performance of this cup and confirms the potential of vitamin E blended HXLPE.
| Original language | English |
|---|---|
| Pages (from-to) | 615-621 |
| Number of pages | 7 |
| Journal | The Journal of Arthroplasty |
| Volume | 30 |
| Issue number | 4 |
| DOIs | |
| Publication status | Published - 1 Apr 2015 |
Keywords
- Femoral head penetration
- Highly-crosslinked polyethylene
- Iso-elastic
- Total hip arthroplasty
- Uncemented monoblock cup
- Vitamin E blended
Fingerprint
Dive into the research topics of 'An uncemented iso-elastic monoblock acetabular component: Preliminary results'. Together they form a unique fingerprint.Cite this