T1 - Allura clarity radiation dose-reduction technology in the hybrid operating room during endovascular aneurysm repair
AU - De Ruiter, Quirina M B
AU - Moll, Frans L.
AU - Gijsberts, Crystel M.
AU - Van Herwaarden, Joost A.
PY - 2016/2/1
Y1 - 2016/2/1
N2 - Purpose: To evaluate the effect of radiation dose reduction with the Allura ClarityIQ image processing technology for fixed C-arms in comparison with a mobile C-arm and an Allura fixed C-arm without ClarityIQ technology during endovascular aneurysm repair (EVAR) procedures. Methods: Radiation dose data from 85 patients (mean age 74.2±7.8 years; 68 men) undergoing EVAR with mobile and fixed C-arm fluoroscopy were retrospectively analyzed. The radiation dose parameters included the kerma area product (KAP), fluoroscopic time (FT), and number of digital subtraction angiography (DSA) frames (FrDSA). KAPtotal consisted of KAPfluoro (KAP for fluoroscopic imaging) and KAPDSA (KAP for DSA and single shots). Linear regression analysis was used to explore differences in the association of KAP with the FT, FrDSA, and body mass index (BMI) among the 3 C-arms. Results: The mean KAPtotal values for mobile, Allura C-arm, and AlluraClarity C-arm for noncomplex EVARs were 56±39, 245±142, and 157±120 Gycm2 (p2 (pfluoro tripled when the mobile C-arm was replaced by the fixed C-arm. There were no significant differences in the KAPfluoro adjusted for the FT between Allura and AlluraClarity (p=0.69). However, there was a major 61% reduction in KAPDSA from 1.36 Gycm2 per DSA frame for Allura to 0.54 Gycm2 per DSA frame with AlluraClarity (p=0.03). For the mobile C-arm, BMI was not associated with KAP (p=0.13). The associations of BMI with KAPfluoro and KAPDSA were significant for both fixed C-arms but were more robust for Allura compared to AlluraClarity (p=0.02 for KAPfluoro and pDSA). Conclusion: Changing a mobile C-arm for a fixed C-arm in a hybrid operating suite increased the average intraoperative dose during EVAR. Upgrading the Allura fixed C-arm with ClarityIQ technology resulted in a 61% reduction in the radiation per DSA frame.
AB - Purpose: To evaluate the effect of radiation dose reduction with the Allura ClarityIQ image processing technology for fixed C-arms in comparison with a mobile C-arm and an Allura fixed C-arm without ClarityIQ technology during endovascular aneurysm repair (EVAR) procedures. Methods: Radiation dose data from 85 patients (mean age 74.2±7.8 years; 68 men) undergoing EVAR with mobile and fixed C-arm fluoroscopy were retrospectively analyzed. The radiation dose parameters included the kerma area product (KAP), fluoroscopic time (FT), and number of digital subtraction angiography (DSA) frames (FrDSA). KAPtotal consisted of KAPfluoro (KAP for fluoroscopic imaging) and KAPDSA (KAP for DSA and single shots). Linear regression analysis was used to explore differences in the association of KAP with the FT, FrDSA, and body mass index (BMI) among the 3 C-arms. Results: The mean KAPtotal values for mobile, Allura C-arm, and AlluraClarity C-arm for noncomplex EVARs were 56±39, 245±142, and 157±120 Gycm2 (p2 (pfluoro tripled when the mobile C-arm was replaced by the fixed C-arm. There were no significant differences in the KAPfluoro adjusted for the FT between Allura and AlluraClarity (p=0.69). However, there was a major 61% reduction in KAPDSA from 1.36 Gycm2 per DSA frame for Allura to 0.54 Gycm2 per DSA frame with AlluraClarity (p=0.03). For the mobile C-arm, BMI was not associated with KAP (p=0.13). The associations of BMI with KAPfluoro and KAPDSA were significant for both fixed C-arms but were more robust for Allura compared to AlluraClarity (p=0.02 for KAPfluoro and pDSA). Conclusion: Changing a mobile C-arm for a fixed C-arm in a hybrid operating suite increased the average intraoperative dose during EVAR. Upgrading the Allura fixed C-arm with ClarityIQ technology resulted in a 61% reduction in the radiation per DSA frame.
KW - Abdominal aortic aneurysm
KW - C-arm image intensifier
KW - Digital subtraction angiography
KW - Dose reduction
KW - Endovascular aneurysm repair
KW - Fluoroscopy
KW - Imaging technology
KW - Kerma area product
KW - Radiation dose
KW - Radiation physics
UR - http://www.scopus.com/inward/record.url?scp=84957811452&partnerID=8YFLogxK
U2 - 10.1177/1526602815622433
DO - 10.1177/1526602815622433
M3 - Article
C2 - 26763260
AN - SCOPUS:84957811452
SN - 1526-6028
VL - 23
SP - 130
EP - 138
JO - Journal of Endovascular Therapy
JF - Journal of Endovascular Therapy
IS - 1
ER -