Abstract
A quality assurance (QA) procedure for magnetic resonance-guided focused ultrasound (MRgFUS) body systems was developed and tested at five institutions. The QA procedure is based on a custom-designed phantom with tissue-mimicking acoustic, MR, and thermal properties specific for MRgFUS. The data acquisition includes assessments of maximum temperature rise, targeting error, signal-to-noise ratio (SNR), and three-dimensional thermal ablation spot size, which are to be tracked over longitudinal repetitions of the QA procedure. The procedure is appropriate for intermediate-level QA, for example, every 20 patients or 6 months, and can be completed in approximately two hours. A subset of the full procedure is recommended for system verification either daily or before each patient (in addition to the vendor-recommended procedure). The phantom and QA procedure have been tested on a variety of custom and commercially available MRgFUS body systems, including systems with 1.5 and 3.0 T static magnetic fields. Normal variability in measurements based on data from five functional MRgFUS body systems over periods of at least 6 months is reported and establishes quantitative criteria for passing the QA test. This QA procedure may be used to increase confidence in the safety and performance of MRgFUS treatments.
| Original language | English |
|---|---|
| Article number | e70585 |
| Journal | Medical physics |
| Volume | 53 |
| Issue number | 8 |
| DOIs | |
| Publication status | Published - Aug 2026 |
Keywords
- focused ultrasound
- FUS
- HIFU
- magnetic resonance imaging
- medical physicist
- MR thermometry
- MR-guided
- MR-HIFU
- MRgFUS
- quality assurance (QA)
- technical guidelines
- thermal ablation
- thermal dosimetry
Fingerprint
Dive into the research topics of 'AAPM Task Group 333 Report: Quality assurance for MRI-guided focused ultrasound body systems'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver