AAPM Task Group 264: The safe clinical implementation of MLC tracking in radiotherapy

Paul J. Keall*, Amit Sawant, Ross I. Berbeco, Jeremy T. Booth, Byungchul Cho, Laura I. Cerviño, Eileen Cirino, Sonja Dieterich, Martin F. Fast, Peter B. Greer, Per Munck af Rosenschöld, Parag J. Parikh, Per Rugaard Poulsen, Lakshmi Santanam, George W. Sherouse, Jie Shi, Sotirios Stathakis

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

3 Citations (Scopus)

Abstract

The era of real-time radiotherapy is upon us. Robotic and gimbaled linac tracking are clinically established technologies with the clinical realization of couch tracking in development. Multileaf collimators (MLCs) are a standard equipment for most cancer radiotherapy systems, and therefore MLC tracking is a potentially widely available technology. MLC tracking has been the subject of theoretical and experimental research for decades and was first implemented for patient treatments in 2013. The AAPM Task Group 264 Safe Clinical Implementation of MLC Tracking in Radiotherapy Report was charged to proactively provide the broader radiation oncology community with (a) clinical implementation guidelines including hardware, software, and clinical indications for use, (b) commissioning and quality assurance recommendations based on early user experience, as well as guidelines on Failure Mode and Effects Analysis, and (c) a discussion of potential future developments. The deliverables from this report include: an explanation of MLC tracking and its historical development; terms and definitions relevant to MLC tracking; the clinical benefit of, clinical experience with and clinical implementation guidelines for MLC tracking; quality assurance guidelines, including example quality assurance worksheets; a clinical decision pathway, future outlook and overall recommendations.

Original languageEnglish
Pages (from-to)e44-e64
JournalMedical Physics
Volume48
Issue number5
DOIs
Publication statusPublished - May 2021

Keywords

  • Humans
  • Particle Accelerators
  • Radiation Oncology
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted
  • Robotics

Fingerprint

Dive into the research topics of 'AAPM Task Group 264: The safe clinical implementation of MLC tracking in radiotherapy'. Together they form a unique fingerprint.