TY - JOUR
T1 - A randomized evaluation of the TriGuard™ HDH cerebral embolic protection device to Reduce the Impact of Cerebral Embolic LEsions after TransCatheter Aortic Valve ImplanTation
T2 - the REFLECT I trial
AU - Lansky, Alexandra J
AU - Makkar, Rajendra
AU - Nazif, Tamim
AU - Messé, Steven
AU - Forrest, John
AU - Sharma, Rahul
AU - Schofer, Joachim
AU - Linke, Axel
AU - Brown, David
AU - Dhoble, Abhijeet
AU - Horwitz, Phillip
AU - Zang, Ming
AU - DeMarco, Frederico
AU - Rajagopal, Vivek
AU - Dwyer, Michael G
AU - Zivadinov, Robert
AU - Stella, Pieter
AU - Rovin, Joshua
AU - Parise, Helen
AU - Kodali, Susheel
AU - Baumbach, Andreas
AU - Moses, Jeffrey
N1 - Publisher Copyright:
© 2021 Published on behalf of the European Society of Cardiology. All rights reserved.
PY - 2021/7/14
Y1 - 2021/7/14
N2 - Aims: The REFLECT I trial investigated the safety and effectiveness of the TriGuard™ HDH (TG) cerebral embolic deflection device in patients undergoing transcatheter aortic valve replacement (TAVR). Methods and results: This prospective, multicentre, single-blind, 2:1 randomized (TG vs. no TG) study aimed to enrol up to 375 patients, including up to 90 roll-in patients. The primary combined safety endpoint (VARC-2 defined early safety) at 30 days was compared with a performance goal. The primary efficacy endpoint was a hierarchical composite of (i) all-cause mortality or any stroke at 30 days, (ii) National Institutes of Health Stroke Scale (NIHSS) worsening at 2-5 days or Montreal Cognitive Assessment worsening at 30 days, and (iii) total volume of cerebral ischaemic lesions detected by diffusion-weighted magnetic resonance imaging at 2-5 days. Cumulative scores were compared between treatment groups using the Finkelstein-Schoenfeld method. A total of 258 of the planned, 375 patients (68.8%) were enrolled (54 roll-in and 204 randomized). The primary safety outcome was met compared with the performance goal (21.8% vs. 35%, P < 0.0001). The primary hierarchical efficacy endpoint was not met (mean efficacy score, higher is better: -5.3 ± 99.8 TG vs. 11.8 ± 96.4 control, P = 0.31). Covert central nervous system injury was numerically lower with TG both in-hospital (46.1% vs. 60.3%, P = 0.0698) and at 5 days (61.7 vs. 76.2%, P = 0.054) compared with controls. Conclusion: REFLECT I demonstrated that TG cerebral protection during TAVR was safe in comparison with historical TAVR data but did not meet the predefined effectiveness endpoint compared with unprotected TAVR controls.
AB - Aims: The REFLECT I trial investigated the safety and effectiveness of the TriGuard™ HDH (TG) cerebral embolic deflection device in patients undergoing transcatheter aortic valve replacement (TAVR). Methods and results: This prospective, multicentre, single-blind, 2:1 randomized (TG vs. no TG) study aimed to enrol up to 375 patients, including up to 90 roll-in patients. The primary combined safety endpoint (VARC-2 defined early safety) at 30 days was compared with a performance goal. The primary efficacy endpoint was a hierarchical composite of (i) all-cause mortality or any stroke at 30 days, (ii) National Institutes of Health Stroke Scale (NIHSS) worsening at 2-5 days or Montreal Cognitive Assessment worsening at 30 days, and (iii) total volume of cerebral ischaemic lesions detected by diffusion-weighted magnetic resonance imaging at 2-5 days. Cumulative scores were compared between treatment groups using the Finkelstein-Schoenfeld method. A total of 258 of the planned, 375 patients (68.8%) were enrolled (54 roll-in and 204 randomized). The primary safety outcome was met compared with the performance goal (21.8% vs. 35%, P < 0.0001). The primary hierarchical efficacy endpoint was not met (mean efficacy score, higher is better: -5.3 ± 99.8 TG vs. 11.8 ± 96.4 control, P = 0.31). Covert central nervous system injury was numerically lower with TG both in-hospital (46.1% vs. 60.3%, P = 0.0698) and at 5 days (61.7 vs. 76.2%, P = 0.054) compared with controls. Conclusion: REFLECT I demonstrated that TG cerebral protection during TAVR was safe in comparison with historical TAVR data but did not meet the predefined effectiveness endpoint compared with unprotected TAVR controls.
KW - Cerebral ischaemia
KW - Diffusion-weighted imaging
KW - Neuroprotection
KW - Stroke prevention
KW - Transcatheter aortic valve replacement
U2 - 10.1093/eurheartj/ehab213
DO - 10.1093/eurheartj/ehab213
M3 - Article
C2 - 34000004
SN - 0195-668X
VL - 42
SP - 2670
EP - 2679
JO - European heart journal
JF - European heart journal
IS - 27